Recall Z-2610-2026
Medline Industries, LP · initiated May 28, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Medline Convenience kits containing recalled Swan-Ganz Catheters CV ANESTHESIA - ROOM SET UP DYNJ905503D DYNJ905503F DYNJ905503G
Codes / lots: DYNJ905503D UDI-DI 10193489430295 lots 21DBG611 21EBI605 21FBN751 DYNJ905503F UDI-DI 10193489967272 lots 21HBL060 21IBW859 22DBC340 22DBS437 22FBR267 22GBR286 22HBS560 22IBA577 22JBI883 22KBD587 22LBF425 23BBN903 23CBL450 23CBW776 23FBC209 23IBU343 24DBQ181 24FBT062 24GBR740 24HBI428 24JBP984 DYNJ905503G UDI-DI 10198459200397 lots 25ABJ069 25ABL749 25CBB762 25DBG799 25EBM994 25GBW226 25LBA950 25LBT019 26ABH519 26ABT315 26BBQ338 26DBJ110
Reason for recall
Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.
FDA-determined cause: Process change control
Action taken
On May 28, 2026, the firm notified customers via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were referred to BD's recall notice, which provides advice for users and instructions to mitigate hub breakage issues. Product return is not required. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.
Details
- Recall number
- Z-2610-2026
- Initiated
- May 28, 2026
- Posted by FDA
- June 26, 2026
- Product code
- CAI – Circuit, Breathing (W Connector, Adaptor, Y Piece)
- Quantity
- 9,272 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.
- Recalling firm
- Medline Industries, LP, Northfield, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2606-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2607-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters CARDIAC MAJOR…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2608-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters CENTRAL AND ARTERIAL…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2609-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters KIT GERMANTOWN CABG…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2611-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2612-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |
| Z-2613-2026 Class II | Medline Convenience kits containing recalled Swan-Ganz Catheters CABG A CDS CDS982269K…Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of… | 05/28/2026Open, Classified |