Recall Z-2613-2025

Medline Industries, LP · initiated July 31, 2025

Class I Open, Classified

Class I – dangerous or defective products that could cause serious health problems or death

Product

Reprocessed ViewFlex Xtra ICE Catheter to visualize cardiac structures; blood flow and other devices within the heart. Medline Item Numbers D087031RH

Codes / lots: Item Number/UDI-DI D087031RH 10197344019557 LOTS EP240731 EP240814 EP240826 EP240903 EP240916 EP241003 EP241014 EP241028 EP241104 EP241112 EP241202 EP241209A EP241219 EP241224 EP250114 EP250116 EP250204 EP250219

Reason for recall

Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual particulates. If devices with residual particulates on patient-contacting surfaces are used, there is a potential risk of an inflammatory response or systemic infection, and if particulate were to dislodge during use and contact circulating blood and/or the heart, it may produce a granulomatous reaction or thrombus formation, which may lead to cerebral or pulmonary embolism, or deep vein thrombosis.

FDA-determined cause: Under Investigation by firm

Action taken

On July 31, 2025, the firm notified customers via first class mail and email of the potential product issue. An updated letter was sent to customers on 8/13/25 with additional item and lot numbers. Affected products are to be returned for credit. Customers were instructed to immediately identify and quarantine all affected product within their possession. Once they have completed and returned the response form listing their affected inventory, they will receive return labels and RGA information. Customers will receive credit when the returned product is received.

Details

Recall number
Z-2613-2025
Initiated
July 31, 2025
Posted by FDA
September 26, 2025
Product code
OWQ – Reprocessed Intravascular Ultrasound Catheter
Quantity
7913 eaches
Distribution
Domestic US distribution nationwide. International distribution pending. No international distribution.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2610-2025 Class IReprocessed Livewire Steerable Electrophysiology Catheter Item Numbers 401575RH 401582RH…Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual… 07/31/2025Open, Classified
Z-2611-2025 Class IReprocessed Webster CS Catheter 10 Electrodes for electrophysiological mapping of…Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual… 07/31/2025Open, Classified
Z-2612-2025 Class IReprocessed Inquiry Steerable Diagnostics Catheter for electrogram recording and cardiac…Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual… 07/31/2025Open, Classified
Z-2614-2025 Class IReprocessed ACUSON AcuNav Ultrasound Catheter for intracardiac and intra-luminal…Specific lots of reprocessed electrophysiology and ultrasound catheters may contain small residual… 07/31/2025Open, Classified