Recall Z-2624-2023
Ortho-Clinical Diagnostics, INc. · initiated August 5, 2023
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
VITROS XT Chemistry Products ALKP Slides
Codes / lots: Product Code: 1053180 UDI:(10758750008346) GEN #21; Expiry Date Range 01 Jul 2023 - 01 June 2024 GEN #23; Expiry Date Range 01 Oct 2023 - 01 Sep 2024 GEN #26; Expiry Date Range 01 Aug 2023 - 01 Dec 2024 GEN #28; Expiry Date Range 01 Aug 2024 GEN #29; Expiry Date Range 01 Sep 2024 - 01 Dec 2024 GEN #30; Expiry Date Range 02 Jan 2025
Reason for recall
Samples containing high levels of ALKP (or an interfering substance) could deplete the available substrate in the reagent layer of a VITROS ALKP Slide and not post the substrate deleption (DP) code, which may result in the System reporting an incorrect value (lower than the actual value), which may result in missed or delayed diagnosis of liver disease.
FDA-determined cause: Under Investigation by firm
Action taken
An "IMPORTANT PRODUCT CORRECTION NOTIFICATION" letter dated 08/17/2023 was issued to impacted consignees. Consignees are provided instructions to utilize when reviewing sample results and any result between 600-1500 U/L should be diluted using a 10x dilution and retested. A confirmation of receipt of the recall letter should be completed and returned by August 25, 2023. Please forward this notification if the affected product was distributed outside your facility. Save this notification with your documentation or post this notification by each VITROS 250/350/FS 5,1/4600/5600/XT 3400/XT 7600 System until the issue has been resolved. If your laboratory has experienced this issue with the product and has not already done so, please report the occurrence to your local Ortho Care Technical Solutions Center. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Details
- Recall number
- Z-2624-2023
- Initiated
- August 5, 2023
- Posted by FDA
- September 21, 2023
- Product code
- CJE – Nitrophenylphosphate, Alkaline Phosphatase Or Isoenzymes
- Quantity
- 402,966 units
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) distribution to countries of: Bermuda, Belgium, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, Italy, India, Japan, Mexico, Netherlands, Norway, Singapore, Spain, Sweden, Portugal, Russia, United Kingdom
- Recalling firm
- Ortho-Clinical Diagnostics, INc., Rochester, NY
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2623-2023 Class II | VITROS XT Chemistry Products TBIL-ALKP SlidesSamples containing high levels of ALKP (or an interfering substance) could deplete the available substrate… | 08/05/2023Open, Classified |