Recall Z-2628-2024

MEDLINE INDUSTRIES, LP - Northfield · initiated April 8, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

See RES description

Codes / lots: 00-398650AJ, Lot Number 23ABN193 ; 00-398650AJ, Lot Number 23ABN291 ; 00-398650AJ, Lot Number 23EBQ823 ; 00-398650AJ, Lot Number 23FBJ579 ; 00-398650AJ, Lot Number 23FBT136 ; 00-398650AJ, Lot Number 23IBB919 ; 00-398650AJ, Lot Number 23KBE097 ; 00-398650AJ, Lot Number 23LBS258 ; 00-398725AL, Lot Number 19BBA371 ; 00-398725AL, Lot Number 19CBE645 ; 00-398725AL, Lot Number 19DBR240 ; 00-398725AL, Lot Number 19EBK403 ; 00-398725AL, Lot Number 19FBU700 ; 00-398725AL, Lot Number 19GBR148 ; 00-398725AN, Lot Number 21FBS869 ; 00-398725AN, Lot Number 21JBN432 ; 00-398725AN, Lot Number 21LBK643 ; 00-398725AN, Lot Number 22ABX426 ; 00-398725AN, Lot Number 22BBR469 ; 00-398725AN, Lot Number 22CBU793 ; 00-398725AN, Lot Number 22OBM117 ; 00-398725AN, Lot Number 22EBK111 ; 00-398725AN, Lot Number 22GBJ289 ; 00-398725AN, Lot Number 22IBU378 ; 00-398725AN, Lot Number 22JBV752 ; 00-398725AN, Lot Number 22LBE100 ; 00-398725AN, Lot Number 23ABO369 ; 00-398725AN, Lot Number 23BBT199 ; 00-398725AN, Lot Number 23DBI838 ; 00-398725AN, Lot Number 23EBN543 ; 00-398725AN, Lot Number 23FBJ242 ; 00-398725AN, Lot Number 23HBB887 ; 00-398725AN, Lot Number 23IBG737 ; 00-398725AN, Lot Number 23JBD877…

Reason for recall

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.

FDA-determined cause: Nonconforming Material/Component

Action taken

Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.

Details

Recall number
Z-2628-2024
Initiated
April 8, 2024
Posted by FDA
August 27, 2024
Product code
OES – Cardiac Catheterization Kit
Quantity
1,679,067 units in total
Distribution
Worldwide distribution.
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2623-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) LOCAL PACK…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2624-2024 Class IIsee RES for descriptionMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2625-2024 Class IISee RES DescriptionMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2626-2024 Class IISee RES descriptionMedical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2627-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) ANGIOGRAPHY…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2629-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) CVC INSERT…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2630-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) DIALYSIS CATH…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2631-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) CVC DOUBLE…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2632-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: 1) NEURO ANGIO…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2633-2024 Class IIMedline procedure packs containing Shenli syringes, labeled as follows: NEURO ANGIO…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2634-2024 Class IIMedline convenience kits, labeled as: 1) PAIN SERVICES CDS , Pack Number CDS985445B ; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified
Z-2635-2024 Class IIMedline Convenience kits, labeled as: 1) NERVE STIMULATOR, Pack Number CDS981579O ; 2)…Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024.… 04/08/2024Open, Classified