Recall Z-2635-2010
Ethicon Endo-Surgery Inc · initiated July 9, 2010
Product
Enseal TRIO Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05
Codes / lots: Product Code ETRIO335H, Lot Code: G4TC14, Exp. Date 2012/05
Reason for recall
The weld that connects the electrode to the active rod was separated. This connection is critical to the proper function of these devices. If the weld is separated, the product will not deliver energy to the jaws, resulting in the potential for a cut/transection without a seal. Ethicon Endo-Surgery identified this issue during a routine finished goods quality inspection.
FDA-determined cause: Other
Action taken
Ethicon Endo-Surgery issued an Urgent: Device Recall, Event 2193 letter dated July 19, 2010 identifying the affected device, the issue prompting the recall, and actions to be taken by consignees. Consignees were instructed: Do not use the affected product. Identify and quarantine affected product from inventory. Complete the Business Reply form and fax to Stericycle at 877-597-9580 within three days To return product, place it in a box with a copy of the Business Reply form and mail to Stericycle using the pre-printed UPS shipping label. Replacement product should be received by the customer within 60 days of return of affected product. Customers can contact Stericycle about this recall at 877-243-7314 and referencing Event 2193.
Details
- Recall number
- Z-2635-2010
- Initiated
- July 9, 2010
- Posted by FDA
- September 30, 2010
- Terminated
- February 14, 2011
- Product code
- NLM – Laparoscope, General & Plastic Surgery, Reprocessed
- Quantity
- 396 instruments
- Distribution
- Nationwide Distribution to the following states: AZ, AL, AZ, CA, CO, CT, FL, GA, IN, KY, MA, MD, MI, MO, MS, NC, NV, NY, OH, PA, TN, TX, WA, and WV.
- Recalling firm
- Ethicon Endo-Surgery Inc, Cincinnati, OH