Recall Z-2637-2017

Beckman Coulter Inc. · initiated May 9, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Access System-Free T4 Assay, B01902E, Catalog No. 33880

Codes / lots: all lots

Reason for recall

The Access immunoassays are susceptible to biotin interference.

FDA-determined cause: Device Design

Action taken

An urgent Medial Device Recall letter dated 5/9/17 was sent to customers to inform them that Beckman Coulter is initiating a field action for Access Total T3, Access Thyroglobulin, Access Free T4, and Access GI Monitor. Beckman Coulter has determined through customer feedback and an internal investigation that the four Access immunoassays listed above are susceptible to biotin interference. During interference testing the interference occurred with samples that contained 100 ng/mL of biotin. This level of biotin is greater than the maximum biotin concentration observed in the normal healthy population. Customers are informed of the impact, action, and resolution. Customers are instructed to respond within 10 days: electronically via email or manually by completing and returning the enclosed response form. Customers with any questions regarding the notice are instructed to contact Customer Technical Support Center: From our website: http://www.beckmancoulter.com By phone: call 1-800-854-3633 in the United States and Canada. Outside the United States and Canada, contact your local Beckman Coulter representative.

Details

Recall number
Z-2637-2017
Initiated
May 9, 2017
Terminated
March 30, 2018
Product code
CEC – Radioimmunoassay, Free Thyroxine
Quantity
1,154,084 units total (167,035 units in US)
Distribution
Algeria, Angola, Australia, Austria, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Djibouti, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Gibraltar, Greece, Guam, Guatemala, Honduras, Hong Kong, Hungary, India, Indonesia…
Recalling firm
Beckman Coulter Inc., Brea, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2635-2017 Class IIAccess Total T3, A34435H, Catalog No. 33830The Access immunoassays are susceptible to biotin interference. 05/09/2017Terminated
Z-2636-2017 Class IIAccess Thyroglobulin, A34085G, Catalog No. 33860The Access immunoassays are susceptible to biotin interference. 05/09/2017Terminated
Z-2638-2017 Class IIAccess System-GI Monitor (CA19-9), A83873D, Catalog No. 387687The Access immunoassays are susceptible to biotin interference. 05/09/2017Terminated