Recall Z-2639-2024
Microtek Medical Inc. · initiated June 3, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
C-Flow Bag Decanter, REF 2000S. Packaged on a form, fill and seal machine with a poly/Tyvek material then placed 50 products to a case (cs).
Codes / lots: SKU/UDI-Device Identifier: 2000S/00748426002085; Lot Numbers: 190220,200424,200424,191214,191025,191118,211024,210602,211016,211016,210912,211024,210912,210912,210912,210727,210828,210608,210608,210411,210411,210323,210118,210113,210113,201123,201009,200423,190429,200619,190719,191025,190614,190715,190614,200424,190614,200619,200914,191225,191027,191106,181005,190715,181123,190429,190606,181207,190220,190805,190503,190304,200331,181207,200914,200914,190606,191209,190719,191225,200424,190606,190715,191027,190429,181005,181207,190606,181123,200311,200311,191118,190826,211116,211116,211116,211116,211024,211024,210714,210714,211116,211024,211024,211116,211016,211016,211016,211016,210912,210912,210827,210827,210912,210912,210828,210828,210828,210828,210827,210608,210608,210608,210608,210602,210504,210411,210411,210411,210323,210323,210323,210323,210113,201123,201123,201009,200903,200914,201009,200903,200903,200903,200903,201009,200619,191106,190304,200914,181207,191210,190614,190220,190715,191225,181005,190614,181005,200914,200619,190719,200423,191106,190220,190429,190826,200619,191118,181005,191225,190614,190429,181207,181005,191118,190429,190220,191027,190715,181123,190110,200903,191…
Reason for recall
During sterile barrier testing performed on the decanter product line, the samples were identified to have pin holes and tears in the packaging film.
FDA-determined cause: Device Design
Action taken
Consignees were notified via email and letter on 06/03/2024. The consignees were instructed to "Segregate Recalled Product/immediately examine the inventory for the lots listed above and remove all products from inventory regardless of its location and segregate this product in a secure location. The lot number can be found on the individual package or case label. If the consignee has affected product in their possession, please dispose of the product and packaging in accordance with hospital, administrative and /or local government policy. If the consignee is unable to dispose of the product, please contact the Customer Service team at the information provided below in this letter. The consignee is requested to please complete the enclosed Ecolab Customer Response Form and e-mail it to recall@ecolab.com, whether they have affected product in their possession or not. If they need assistance with filling out the form, please contact the sales representative. If the consignee has further distributed this product, instead complete the enclosed Ecolab Distributor Response Form and e-mail it to recall@ecolab.com whether their customers have affected product in their possession or not. Please identify the consignee's customers and notify recall@ecolab.com so Ecolab may notify them of this Field Recall. It is important that the firm receive the completed Ecolab Customer Response Form or Ecolab Distributor Response Form by June 14, 2024. The consignee is instructed to please contact the Customer Service at 1-800-824-3027 from 8:30 a.m. to 5 p.m. (CST) to obtain a Returned Material Authorization (RMA) number if they are unable to dispose of the affected product, or email at csdomesticorders@ecolab.com or cs-distributorrequests@ecolab.com. The consignee is to package the decanter(s) that are being returned in an appropriate shipping box. Place a copy of the response form in the box, appropriately seal the box, and send it on or before June 14, 2024, to: ATTN: Belissa del Val
Details
- Recall number
- Z-2639-2024
- Initiated
- June 3, 2024
- Posted by FDA
- August 16, 2024
- Product code
- KPE – Container, I.V.
- Quantity
- 10,366 cases (518,300 units)
- Distribution
- Distributed nationwide and internationally to Canada, Mexico, New Zealand, Belize, and Hong Kong.
- Recalling firm
- Microtek Medical Inc., Saint Paul, MN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2640-2024 Class II | Bag Decanter II, REF 2002S. Packaged on a form, fill and seal machine with a poly/Tyvek…During sterile barrier testing performed on the decanter product line, the samples were identified to have… | 06/03/2024Open, Classified |
| Z-2641-2024 Class II | Bottle Decanter, REF 2004S. Packaged on a form, fill and seal machine with a poly/Tyvek…During sterile barrier testing performed on the decanter product line, the samples were identified to have… | 06/03/2024Open, Classified |
| Z-2642-2024 Class II | Vial Decanter, REF 2006S. Packaged on a form, fill and seal machine with a poly/Tyvek…During sterile barrier testing performed on the decanter product line, the samples were identified to have… | 06/03/2024Open, Classified |
| Z-2643-2024 Class II | Transfer Device, REF 2008S. Packaged on a form, fill and seal machine with a poly/Tyvek…During sterile barrier testing performed on the decanter product line, the samples were identified to have… | 06/03/2024Open, Classified |