Recall Z-2644-2024
Ventana Medical Systems, Inc. · initiated June 27, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody, REF: 790-4506, IVD
Codes / lots: UDI: 04015630972579/J04613, J11853, J17541, J25047, J30286, K00982, K06239, K09880, K14266, K19784, K26461, and M00669
Reason for recall
Their is a potential for high background and off-target staining when using the monoclonal primary antibody which can result in false positive results.
FDA-determined cause: Unknown/Undetermined by firm
Action taken
On July 5, 2024, Ventana Medical Systems, Inc. a subsidiary of Roche Diagnostics issued a "Urgent Medical Device Correction" via UPS to affected consignees. Ventana asked consignees to take the following actions: 1. Discontinue use and discard any remaining affected product in your inventory according to your local waste guidelines. 2. Please consult with the physician or pathologist at your facility to determine specific clinical implications for your patients. 3. Review all positive results generated with the affected lots. Follow your standard laboratory operating procedures to investigate any suspected false positive results generated with the affected lots. 4. If you performed your initial validation using an affected lot, you should revalidate. 5. Refer to the Product Replacement or Credit section on page three of this UMDC. To receive replacement product OR credit for the unexpired product currently in your inventory, use the following link to access the form: https://go.roche.com/TP-02194. 6. Complete all sections of the enclosed Recall Response form (TP-02195); fax or email it according to the instructions on the form even if you are not requesting replacement product or credit. " Share this notification with all device users within your facility and network to ensure they are aware of this recall. This recall notice should also be shared with any organization where the potentially affected devices have been transferred. " File this Urgent Medical Device Correction (UMDC) for future reference.
Details
- Recall number
- Z-2644-2024
- Initiated
- June 27, 2024
- Posted by FDA
- August 16, 2024
- Product code
- NJT – Immunohistochemistry Reagents And Kits
- Quantity
- 22,839 kits
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland…
- Recalling firm
- Ventana Medical Systems, Inc., Oro Valley, AZ