Recall Z-2673-2011

Instrumentation Laboratory Co. · initiated May 11, 2011

Terminated

Product

Instrumentation Laboratory GEM Premier 3500 Blood Gas Analyzer PN 00026000000 The GEM Premier 3500 blood gas analyzer is a portable system for use by health care professionals to rapidly analyze whole blood samples at the point of health care delivery in a clinical setting. The instrument provides quantitative measurements of whole blood for pH, pC02, p02, Na+, K+, Ca++, glucose, lactate and hematocrit. These parameters, along with derived parameters, aid in the diagnosis of a patient's acid/base status, oxygen delivery capacity, and electrolyte and metabolite balance.

Codes / lots: Serial Numbers range: 09020100 through 11041157.

Reason for recall

Software upgrade: misuse of the barcode gun by repeatedly clicking can cause the instrument to attach test results to an incorrect patient ID.

FDA-determined cause: Software design

Action taken

Instrumentation Laboratories sent a "FIELD SAFETY NOTIFICATION" letter dated April 2011 to all affected customers. The letter described the product, problem, and the actions to be taken by the customers. Included with the letter was an upgrade package containing Software V7.2.2 disk along with installation instructions. Also included with the upgrade kit was a Upgrade Tracking Form to be completed and returned via fax at 781-861-4207 or via e-mail at ra-usa@ilww.com. Customers with questions were directed to contact their local representative.

Details

Recall number
Z-2673-2011
Initiated
May 11, 2011
Posted by FDA
June 27, 2011
Terminated
March 15, 2013
Product code
CHL – Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Quantity
840 units
Distribution
Worldwide Distribution-- USA (nationwide) and the countries of AUSTRALIA, AUSTRIA, BELGIUM, CANADA, CHINA, CZECH REPUBLIC, DENMARK, EGYPT, FINLAND, FRANCE, GERMANY, GREECE , INDONESIA, ISRAEL, ITALY, JAPAN , KOREA, REPUBLIC OF CHINA, MALAYSIA, NETHERLANDS , PHILIPPINES, POLAND, PORTUGAL, SPAIN, SWEDEN, SWITZERLAND, THAILAND ,TURKEY, and UNITED KINGDOM.
Recalling firm
Instrumentation Laboratory Co., Bedford, MA

Official FDA recall record