Recall Z-2673-2026

Zimmer Surgical Inc · initiated June 9, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Brand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems Dual Hose with CPC Connectors Model/Catalog Number: 60-1812-101-00 Software Version: N/A Product Description: Dual Sterile Extension Hose with PLC Connector Component: N/A

Codes / lots: Lot Code: Model No. 60-1812-101-00 ; Case (20-pack) UDI (01)00889024381162(17)271125(10)78938271 ; Box (10-Pack) UDI (01)00889024378469(17)271125(10)78938271 ; Individual UDI (01)00889024367173(17)271125(10)78938271 ; Lot Number 78938271 Model No. 60-1812-101-00 ; Case (20-pack) UDI (01)00889024381162(17)271126(10)78938272 ; Box (10-Pack) UDI (01)00889024378469(17)271126(10)78938272 ; Individual UDI (01)00889024367173(17)271126(10)78938272 ; Lot Number 78938272 Model No. 60-1812-101-00 ; Case (20-pack) UDI (01)00889024381162(17)271127(10)78938273 ; Box (10-Pack) UDI (01)00889024378469(17)271127(10)78938273 ; Individual UDI (01)00889024367173(17)271127(10)78938273 ; Lot Number 78938273 Model No. 60-1812-101-00 ; Case (20-pack) UDI (01)00889024381162(17)280206(10)80555986 ; Box (10-Pack) UDI (01)00889024378469(17)280206(10)80555986 ; Individual UDI (01)00889024367173(17)280206(10)80555986 ; Lot Number 80555986 Model No. 60-1812-101-00 ; Case (20-pack) UDI (01)00889024381162(17)280102(10)80555987 ; Box (10-Pack) UDI (01)00889024378469(17)280102(10)80555987 ; Individual UDI (01)00889024367173(17)280102(10)80555987 ; Lot Number 80555987 Model No. 60-1812-101-00 ; Case (20-pack) UDI…

Reason for recall

Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized infection. If discovered prior to use, may result in clinically insignificant extension of surgery to find another readily available device.

FDA-determined cause: Under Investigation by firm

Action taken

On June 8, 2026, the firm began notifying consignees titled "URGENT MEDICAL DEVICE RECALL." Customers were instructed to locate and quarantine affected product in their inventory. Customers may either return the affected product, or a Zimmer Biomet sales representative may remove and return the affected product on the customer's behalf. If the product has been further distributed, the distributor should notify downstream customers and accounts of the recall. All customers should ensure that affected personnel are aware of the recall. This recall is related to prior recall RES 98665, conducted to remove Mixing Bowl and Spatula due to similar issues with seal integrity.

Details

Recall number
Z-2673-2026
Initiated
June 9, 2026
Posted by FDA
July 2, 2026
Product code
KCY – Tourniquet, Pneumatic
Quantity
45160
Distribution
Worldwide distribution - US Nationwide and the countries of Canada and EMEA.
Recalling firm
Zimmer Surgical Inc, Dover, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2674-2026 Class IIBrand Name: Sterile Tourniquet Hoses Product Name: A.T.S 4000 TS Tourniquet Systems…Potential sterility breach in sealing area of sterile pouch. If used, issue may result in moderate localized… 06/09/2026Open, Classified