Recall Z-2728-2016

Hospira Inc. · initiated February 6, 2013

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Plum A+3 Infusion Pump System. List Number 12348. The Plum A+ Infusion Systems are a multi-purpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self-contained, microprocessor based infusion devices.

Codes / lots: List number 12348, Serial Numbers: 13740156, 13740256, 13740266, 13740351, 13740380, 13740451, 13740508, 13740612, 13740632, 13740731, 13740732, 13740733, 13740734, 13740735, 13740736, 13740737, 13740738, 13740739, 13740740, 13740741, 13740742, 13740743, 13740744, 13740745, 13740746, 13740747, 13740748, 13740749, 13740750, 13740751, 13740752, 13740753, 13740754, 13740755, 13740756, 13740757, 13740758, 13740759, 13740760, 13740761, 13740762, 13740763, 13740764, 13740765, 13740766, 13740767, 13740768, 13740769, 13740770, 13740771, 13740772, 13740773, 13740774, 13740775, 13740776, 13740777, 13740778, 13740779, 13740780, 13740781, 13740782, 13740783, 13740784, 13740785, 13740786, 13740787, 13740788, 13740789, 13740790, 13740791, 13740792, 13740793, 13740794, 13740795, 13740796, 13740797, 13740798, 13740799, 13740800, 13740801, 13740802, 13740803, 13740804, 13740805, 13740806, 13740807, 13740808, 13740809, 13740810, 13740811, 13740812, 13740813, 13740814, 13740815, 13740816, 13740818, 13740819, 13740820, 13740821, 13740822, 13740823, 13740824, 13740825, 13740826, 13740827, 13740828, 13740829, 13740830, 13740831, 13740832, 13740833, 13740834, 13740835, 13740836, 13740837, 13740838…

Reason for recall

The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and delay or interruption of therapy.

FDA-determined cause: Nonconforming Material/Component

Action taken

Hospira sent an Urgent Device Field Correction letter, dated February 6, 2013, to customers. Customers were requested to return a reply form to Stericycle indicating that they have received the letter, and notified their users and/or customers. The customers were asked to supply the number of Plum A+ infusers at their facility. Product will be corrected by Hospira in the field.

Details

Recall number
Z-2728-2016
Initiated
February 6, 2013
Posted by FDA
September 6, 2016
Terminated
November 14, 2016
Product code
FRN – Pump, Infusion
Quantity
357,778 total
Distribution
United States nationwide distribution including PR, and USVI. Costa Rica, Turks and Caicos, and Dominican Republic.
Recalling firm
Hospira Inc., Lake Forest, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2722-2016 Class IIPlum A+ Hyperbaric Infusion Pump. List Number 11005. The Plum A+ Infusion Systems are a…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2723-2016 Class IIPlum A+ Infusion Pump. List Number 11971. The Plum A+ Infusion Systems are a…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2724-2016 Class IIPlum A+ Infusion Pump. List Number 11973. The Plum A+ Infusion Systems are a…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2725-2016 Class IIPlum A+3 Infusion Pump. List Number 12391. The Plum A+ Infusion Systems are a…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2726-2016 Class IIPlum A+ Infusion Pump with Hospira MedNet Software. List Number 20679. The Plum A+…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2727-2016 Class IIPlum A+ Infusion Pump with Hospira MedNet Software. List Number 20792. The Plum A+…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2729-2016 Class IIPlum A+3 Infusion Pump. List Number 12618. The Plum A+ Infusion Systems are a…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated
Z-2730-2016 Class IIPlum A+3 Infusion Pump with Hospira MedNet Software. List Number 20678. The Plum A+…The fluid shield diaphragm for the Plum A+ Infusion pumps was undersized, possibly resulting in alarms and… 02/06/2013Terminated