Recall Z-2730-2017

Beckman Coulter Inc · initiated May 15, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Beckman Coulter Synchron AQUA CAL Level 1, Part No. 471288 Product Usage: Used for the system calibration of Na, K, Cl, Urea N, Urea, Glucose Creatine, Ca, CO2, and P in UniCel DxC Systems.

Codes / lots: Lot No. 72645074, Exp. Date 30 Apr 2018

Reason for recall

The firm identified a negative shift of 5mmol/L for sodium recovery in Synchron Control Level 3 when calibrated with AQUA Cal 1.

FDA-determined cause: Nonconforming Material/Component

Action taken

Recall notices will be sent via mail or email to customers the week of 05/15/2017. Customers are instructed to discard the affected calibrator and request a replacement. Call Client Services at (800) 526-3821 (US only)

Details

Recall number
Z-2730-2017
Initiated
May 15, 2017
Posted by FDA
July 13, 2017
Terminated
September 9, 2020
Product code
JIX – Calibrator, Multi-Analyte Mixture
Quantity
4444 AQUA CAL Level 1 kits
Distribution
US Nationwide Distribution in the states of AL, AR, AK, CA, CO, DE, FL, GA, ID, IL, IN, IO, KS, KY, LA, MA, MD, MS, MI, ME, MO, NE, NV, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WV, WI, and Puerto Rico Brunei, Canada, China, Czech Republic, France, Guatemala, Hong Kong, India, Ireland, Italy, Japan, Lebanon, Mexico, Netherlands, Panama, Paraguay, Philippines, Qatar, South…
Recalling firm
Beckman Coulter Inc, Brea, CA

Official FDA recall record