Recall Z-2730-2024
Stryker Corporation · initiated July 24, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Stryker iBur 3.0mm Diamond Match Head, Distal Bend - intended to cut bone in the following manner: drilling, reaming, decorticating, shaping, dissecting, shaving, and smoothing for the following medical applications: Neuro; Spine; Ear, Nose, and Throat (ENT)/ Otorhinolaryngology; and Endoscopic applications Ref: 8431-107-030D
Codes / lots: GTIN: 07613327509229 Lot Numbers: 22334017 23047017 23095017 23198017
Reason for recall
Potential for the product to exhibit temperatures higher than specification where the bur shank meets the distal bushing. Temperatures higher than specification may lead to the potential for minor tissue/structure damage or tissue/structure damage from thermal injury which may need medical/surgical intervention.
FDA-determined cause: Under Investigation by firm
Action taken
Stryker issued Urgent Medical Device Recall letter on 7/24/424 to Materials Manager, Risk Manager, OR Director via Fed'x. Letter states reason for recall, health risk, and action to take: 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email Instruments.Recalls@stryker.com. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A credit will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. If you have any questions about this matter, contact Stryker Instruments by phone (269) 568-1481 or email at Instruments.Recalls@stryker.com.
Details
- Recall number
- Z-2730-2024
- Initiated
- July 24, 2024
- Posted by FDA
- August 26, 2024
- Product code
- HBE – Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Quantity
- 2507 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, France, Hong Kong, India, Italy, Japan, Japan, Netherlands, South Korea, Spain, Sweden, Taiwan, United Kingdom.
- Recalling firm
- Stryker Corporation, Portage, MI
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2722-2024 Class II | Stryker iBur 3.0mm Precision Round, Distal Bend- Intended to cut bone in the following…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2723-2024 Class II | Stryker iBur 4.0mm Precision Round, Distal Bend- intended to cut bone in the following…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2724-2024 Class II | Stryker iBur 2.0mm Diamond Round, Distal Bend- intended to cut bone in the following…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2725-2024 Class II | Stryker iBur 3.0mm Diamond Round, Distal Bend - intended to cut bone in the following…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2726-2024 Class II | Stryker iBur 4.0mm Diamond Round, Distal Bend- intended to cut bone in the following…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2727-2024 Class II | Stryker iBur 3.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2728-2024 Class II | Stryker iBur 4.0mm Coarse Diamond Round, Distal Bend - intended to cut bone in the…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2729-2024 Class II | Stryker iBur 5.0mm Coarse Diamond Round, Distal Bend intended to cut bone in the…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |
| Z-2731-2024 Class II | Stryker iBur 3.0mm Precision Match Head, Distal Bend- intended to cut bone in the…Potential for the product to exhibit temperatures higher than specification where the bur shank meets the… | 07/24/2024Completed |