Recall Z-2736-2018

Hudson Scientific LLC · initiated July 2, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

iTOVi Tracker/Scanner, Model Nos. ITS10 and ITS15 Intended to rank galvanic skin response measurements. The results are used by the software to generate a report that then makes a recommendation on nutritional supplements.

Codes / lots: Device Lot/ Batch Numbers: 012016, 032016, 112016, 122016, 012017, 022017, 032017, 092017, 112017, 012018 Device Serial Numbers: 35,949 MAC Numbers ranging from 0F:72: D9:E9: F4:5A to FF: FF: AB:37: 7A:0C

Reason for recall

The device and its accessories may not have been manufactured according to current Good Manufacturing Practices. In addition, the firm has received three complaints regarding electric shocks received when using the device.

FDA-determined cause: Nonconforming Material/Component

Action taken

On July 2, 2018, the firm notified customers of the recall via an emailed Urgent Medical Device Removal letter. Customers were informed that the company had received three complaints of electrical shock when using the device, and that the device and its accessories may not have been manufactured according to current good manufacturing practices. ACTIONS TO BE TAKEN BY THE CUSTOMER 1. Dispose of the iTOVi Scanner and its power adapter and charger in the trash. 2. Fill out and return to us the enclosed Attachment A: Urgent Removal Response Letter. The bottom of the form identifies three different methods of returning this form to iTOVi. 3. Forward this letter to any users to whom you may have distributed the device. 4. If you or others have encountered any illnesses or injuries during or after use of any iTOVi Tracker/ Scanner products, iTOVi recommends that you arrange for yourself and other product users to seek medical attention. TYPE OF ACTION BY THE COMPANY: As of June 15, 2018, all iTOVi Tracker/ Scanner software applications (apps) will be deactivated via the following steps: 1. Using the server-based software that handles iTOVi account logins, iTOVi will disable the iTOVi computer software application that runs on iOS and Android operating systems for all iTOVi devices. 2. iTOVi will issue an update to the iTOVi software application which displays a message saying that the iTOVi Scanner device is no longer usable and should be disposed of in the trash. iTOVi is in the process of creating a more reliable, innovative iTOVi Scanner, and will be sending out a new device to replace older models.

Details

Recall number
Z-2736-2018
Initiated
July 2, 2018
Terminated
January 13, 2020
Product code
GZO – Device, Galvanic Skin Response Measurement
Quantity
35,949 units
Distribution
Nationwide distribution. Worldwide distribution.
Recalling firm
Hudson Scientific LLC, Hudson, MI