Recall Z-2753-2019

Roche Diabetes Care, Inc. · initiated May 3, 2019

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Accu-Chek Connect Diabetes Management App

Codes / lots: Connect App version 2.1.6.3 when used with Android OS 8.0 and above.

Reason for recall

Users with Android OS 8.0 and above may be unable to transfer values obtained on their meter to the App.

FDA-determined cause: Software in the Use Environment

Action taken

On 5/2/19, the firm notified affiliates of the issue via a Quality Notice. On 5/16/19, the firm issued an updated Quality Notice stating that a new version of the software, 2.1.7.2 would soon become available which would resolve the data transfer issues. Call centers were also provided with a workaround for customers who called in reporting an issue with transferring results. On 5/24/19, version 2.1.7.2 of the Connect App was released for Android on the Google Play store. This version of the app corrects the data transfer issue. At this time, users were notified via the "What's New" section of Google Play. These users were informed that the new version addresses issues with wireless data transfers from blood glucose meters. Customers with questions regarding this event should call the ACCU-CHEK Customer Care Software support line at 1-800-628-3346.

Details

Recall number
Z-2753-2019
Initiated
May 3, 2019
Terminated
October 30, 2020
Product code
NDC – Calculator, Drug Dose
Distribution
Nationwide and worldwide usage of the Connect App.
Recalling firm
Roche Diabetes Care, Inc., Indianapolis, IN