Recall Z-2754-2020

Becton Dickinson & Co. · initiated July 20, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD MAX SARS-CoV-2 Reagents, Cat. No. 445003 (UDI 00382904450038) and 445003-01 (UDI (01)603829044500301) - Product Usage: intended for the qualitative detection of nucleic acid from the SARS-CoV-2 in nasopharyngeal, nasal, mid-turbinate, and oropharyngeal swab specimens, nasopharyngeal wash/aspirate or nasal aspirates obtained from individuals suspected of COVID-19 by their healthcare provider.

Codes / lots: All lots are affected

Reason for recall

BD is experiencing elevated rates of potential false positive results from certain customers when using the BD SARS-CoV-2 Reagents for the BD MAX System. False positive test results could lead to a delay in diagnosis and potentially cause the patient to be exposed to COVID-19 if they are isolated with an infected individual.

FDA-determined cause: Under Investigation by firm

Action taken

On July 20, 2020, the firm sent Urgent Medical Device Product Advisory letters to their customers. Customers were informed that all lots are impacted by the elevated rates of potential false positive results. Customers were advised that all positive results are considered presumptive. BD recommends that testing for any positive samples be repeated using an alternate authorized test method. Customers were asked to share the notification with all users of the reagents within their facility to ensure that they are also aware of this product advisory. Customers are to complete and return the customer response form to the recalling firm. US customers who require further assistance may contact BD Customer/Technical Support at 800-638-8663, Monday-Friday, between 7AM-7PM (EST). OUS customers may contact their local BD representative or distributor.

Details

Recall number
Z-2754-2020
Initiated
July 20, 2020
Terminated
July 13, 2021
Product code
QJR – Reagents, 2019-Novel Coronavirus Nucleic Acid
Quantity
30,392 to date
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AZ, CA, CO, DC, FL, GU, HI, IL, IN, KY, MD, MI, MO, NC, NJ, NY, PA, SC, TN, TX, WI, WY and the countries of Argentina, Australia, Belgium, Brazil, Canada, Chile, Colombia, Hong Kong, Japan, Mexico, New Zealand, Peru, Philippines, Singapore.
Recalling firm
Becton Dickinson & Co., Sparks, MD