Recall Z-2756-2017

PerkinElmer Health Sciences, Inc. · initiated May 31, 2017

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

PerkinElmer 226 Sample Collection Device The PerkinElmer 226 Sample Collection Device is intended to be used as a medium to collect and transport whole blood specimen spots to a laboratory in Newborn Screening.

Codes / lots: lot numbers 317113 and 317114

Reason for recall

PerkinElmer 226 Sample Collection Devices Lot Number 317114 had serial numbers printed on the pack labels that did not match the serial number range of the devices contained within the pack.

FDA-determined cause: Packaging

Action taken

PerkinElmer sent an Urgent Medical Device Correction letter dated May 30, 2017, to the sole consignee. PerkinElmer requested the consignee implement the following actions: 1. Discontinue use of the pack label to track the device serial numbers distributed to your sample submitters. 2. Verify submitter of devices received without provider information prior to demographic entry. 3. Verify the serial number of the pack label matches the device serial numbers within the pack of your undistributed inventory prior to distributing to your sample submitter or, 4. Return undistributed inventory to PerkinElmer for inspection , and rework if required. For further questions please call (864) 299-8787.

Details

Recall number
Z-2756-2017
Initiated
May 31, 2017
Posted by FDA
July 24, 2017
Terminated
July 27, 2017
Product code
PJC – Newborn Screening Specimen Collection Paper
Quantity
99,968 units
Distribution
US Distribution to the state of : AL
Recalling firm
PerkinElmer Health Sciences, Inc., Greenville, SC

Official FDA recall record