Recall Z-2757-2026

Boston Scientific Corporation · initiated June 11, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

FARADRIVE STEERABLE SHEATH CLEAR OUS, used to facilitate access to the vasculature and into the chambers of the heart, Material Number M004PFCE21M402, IFU Reference Number 51659727-01

Codes / lots: UDI-DI: 00191506042516; Lots: CL12917, CL13004, CL13017, CL13046, CL13060, CL13080, CL13103, CL13122, CL13123, CL13142, CL13160, CL13196, CL13212, CL13219, CL13231, CL13260, CL13291, CL13340, CL13375, CL13401, CL13422, CL13468, CL13519, CL13573, CL13578, CL13614, CL13646, CL13670, CL13671, CL13708, CL13719, CL13722, CL13732, CL13760, CL13801, CL13802, CL13820, CL13833, CL13856, CL13910, CL13911, CL13925, CL13940, CL13971, CL13972, CL13988, CL14025, CL14039, CL14050, CL14051, CL14076, CL14077, CL14119, CL14120, CL14135, CL14137, CL14171, CL14187, CL14226, CL14249, CL14286, CL14287, CL14345, CL14346, CL14375, CL14376, CL14396, CL14397, CL14439, CL14488, CL14500, CL14501, CL14518, CL14519, CL14581, CL14582, CL14633, CL14635, CL14670, CL14697, CL14711, CL14749, CL14750, CL14769, CL14770, CL14813, CL14814, CL14854, CL14855, CL14870, CL14871, CL14928, CL14929, CL14933, CL14934, CL14935, CL14968, CL14969, CL15005, CL15009, CL15067, CL15107, CL15108, CL15109, CL15128, CL15129, CL15191, CL15192, CL15193, CL15239, CL15240, CL15272, CL15318, CL15358, CL15397, CL15427, CL15474, CL15490, CL15539, CL15568, CL15569, CL15597, CL15598, CL15637, CL15638, CL15640, CL15754, CL15755, CL15775…

Reason for recall

Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are associated with hemostatic valve damage related to dilator insertion orientation during device preparation. Valve damage may result in air ingress, potential air embolism, and associated patient harm. This risk can be mitigated by inserting the dilator straight through the center of the valve and fully advancing it into the sheath body.

FDA-determined cause: Use error

Action taken

An Important Medical Device Product Advisory notification letter dated 6/11/26 was sent to customers. Actions " Follow the revised IFU provided in Appendix 1. " Forward this notice to relevant personnel and any facilities to which the product was transferred. " Complete and return the enclosed Acknowledgement Form. " Report any adverse events or quality concerns associated with use of these devices to Boston Scientific via email at BSCComplaints@bsci.com or FDA s MedWatch Adverse Event Reporting program [www.fda.gov/MedWatch/report.htm or 1.800.FDA.1088 (332.1088)]. If you require additional assistance or more information regarding this communication, please contact your local Boston Scientific representative.

Details

Recall number
Z-2757-2026
Initiated
June 11, 2026
Posted by FDA
July 20, 2026
Product code
DRA – Catheter, Steerable
Quantity
193,000 units
Distribution
Global Distribution. US Nationwide.
Recalling firm
Boston Scientific Corporation, Saint Paul, MN

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2754-2026 Class IIFARADRIVE Steerable Sheath Clear IDE, used to facilitate access to the vasculature and…Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are… 06/11/2026Open, Classified
Z-2755-2026 Class IIFARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access to the vasculature and into…Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are… 06/11/2026Open, Classified
Z-2756-2026 Class IIFARADRIVE STEERABLE SHEATH (CLEAR) - US, used to facilitate access to the vasculature…Boston Scientific's complaint investigations have identified that a subset of air ingress complaints are… 06/11/2026Open, Classified