Recall Z-2772-2017

Sorin Group USA, Inc. · initiated June 30, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

STOCKERT Heater-Cooler System 3T, 240V/60 Hz

Codes / lots: Item No. 16-02-81; Serial No. 16S10743-16S11708

Reason for recall

Plastic heat exchanger fibers separating the blood compartment from the water compartment in certain oxygenators may allow diffusion of hydrogen peroxide in a quantity that exceeds allowable limits.

FDA-determined cause: Device Design

Action taken

"Consignees were sent a recall notification, dated 06/30/2017, via mail. Instructions included a recommendation from LivaNova to use its oxygenators, or other oxygenators that are equipped with compatible heat exchangers, to ensure that the permeability rate is within the prescribed allowable limits (ALs). Depending on whether or not the oxygenator was from LivaNova, there were specific instructions to follow for each type and source. Consignees were also instructed to complete and return the Confirmation Form. If product was further distributed, the firm asks that further customers be notified of the recall and that customers contact the LivaNova Quality Assurance Team at usfsn@livanova.com. Questions regarding the recall can be directed to 800-986-4702 or 3T.US@livanova.com. Any adverse reactions are to be reported to FDA's MedWatch Adverse Event Reporting program."

Details

Recall number
Z-2772-2017
Initiated
June 30, 2017
Posted by FDA
July 25, 2017
Terminated
October 10, 2023
Product code
DWC – Controller, Temperature, Cardiopulmonary Bypass
Quantity
7,546 units
Distribution
"Domestic: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, & WY. Foreign: AL, AE, AR, AT, AU, AZ, BD, BE, BG, BH, BR, BY, CA, CH, CL, CN, CO, CR, CY, CZ, DE, DK, DZ, EC, EE, EG, ES, ET, FI, FR, GB, GE, GR, GY, HK, HR, HU, IE, ID, IL, IN, IQ, IR, IT…
Recalling firm
Sorin Group USA, Inc., Arvada, CO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2771-2017 Class IISTOCKERT Heater-Cooler System 3T, 230 VPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain… 06/30/2017Terminated
Z-2773-2017 Class IISTOCKERT Heater-Cooler System 3T, 208V/60 HzPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain… 06/30/2017Terminated
Z-2774-2017 Class IISTOCKERT Heater-Cooler System 3T, 127V/60 HzPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain… 06/30/2017Terminated
Z-2775-2017 Class IISTOCKERT Heater-Cooler System 3T, 120V/60 HzPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain… 06/30/2017Terminated
Z-2776-2017 Class IISTOCKERT Heater-Cooler System 3T, 200V/50 Hz/60 HzPlastic heat exchanger fibers separating the blood compartment from the water compartment in certain… 06/30/2017Terminated