Recall Z-2788-2026

Medline Industries, LP · initiated June 11, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medline medical convenience kits labeled as: 1) KIT PEDIATRIC OPEN HEART MOSES, Medline Kit SKU DYNJ910103B; 2) BASIC HEART CDS, Medline Kit SKU DYNJ910961A.

Codes / lots: Medline Kit SKU DYNJ910103B, UDI/DI 10198459346873 (each) 40198459346874 (case), Lot Number 25FBR237; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25ABM761; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25BBD107; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBD534; Medline Kit SKU DYNJ910961A, UDI/DI 10198459213342 (each) 40198459213343 (case), Lot Number 25FBL772.

Reason for recall

Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL which may be contaminated with low levels of Bacillus species.

FDA-determined cause: Process change control

Action taken

Medline Industries, LP issued an URGENT MEDICAL DEVICE RECALL notice to its consignees via email and USPS first class mail. The notice explained the issue with the device, potential harms, and requested the following: REQUIRED ACTION: 1. Immediately check stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete the response form. Please list the quantity of affected product in inventory on the form. Even if there is no affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com 3. Upon receipt of the response form, the consignee account will receive over-labels to place on affected inventory, with instructions for the staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who transferred the affected product were directed to notify those to which the product was distributed or transferred. For contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-2788-2026
Initiated
June 11, 2026
Posted by FDA
July 23, 2026
Product code
OEZ – Cardiovascular Procedure Kit
Quantity
155 kits
Distribution
US: CT, FL, KS, OH, OR, NY, MD, MT, VA, MA.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2789-2026 Class IIMedline medical convenience kits labeled as: 1) SHOULDER PACK, Medline Kit SKU…Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL… 06/11/2026Open, Classified
Z-2790-2026 Class IIMedline medical convenience kits labeled as: NEURO PACK, Medline Kit SKU DYNJ59397PMedline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL… 06/11/2026Open, Classified
Z-2791-2026 Class IIMedline medical convenience kits labeled as: 1) TOTAL JOINT PACK Medline Kit SKU…Medline Industries, LP is issuing a recall for specific Medline convenience kits containing BD PurPrep 26 mL… 06/11/2026Open, Classified