Recall Z-2793-2020

Brainlab AG · initiated July 16, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BRAINLAB Kick 2; Model/catalogue (article) numbers for Kick: 18170; 18070 UDI (GTIN) for Kick: 04056481006655

Codes / lots: KICK 2 NAVIGATION STATION, Serial Numbers: 0164217001-18170, S/N of affected NDI Camera: P7-12520 2703119001-18170, S/N of affected NDI Camera:P7-20938 2933119001-18170, S/N of affected NDI Camera:P7-21378 6171619001-18170, S/N of affected NDI Camera:P7-18830 8745417001-18170, S/N of affected NDI Camera:P7-15162 9178819001-18170, S/N of affected NDI Camera:P7-16025 9607118001-18170, S/N of affected NDI Camera:P7-13227 5469918001-18170, S/N of affected NDI Camera: P7-14965 KICK NAVIGATION STATION, Serial Number: 1190017001-18070, S/N of affected NDI Camera: P7-06110

Reason for recall

Incorrect manufacturer calibration

FDA-determined cause: Process control

Action taken

1. Refer to the appendix for the list of Brainlab Curve or Kick Navigation Systems with an affected camera installed to determine if your navigation system is affected. 2. For Cranial or ENT navigation software in combination with an affected system camera installed: Do not use the Z-touch Laser Pointer for patient registration. Use other instruments with the standard instrument pivoting method for Surface Matching or Landmark Registration, or Automatic Image Registration (AIR) if available. 3. Please continue to follow the instructions and warnings as described in the user guide. Especially relevant is the following warning, in addition to the instructions on the registration verification page appearing in the Cranial or ENT software: Verify accuracy at multiple anatomical landmarks, especially in the region of interest, as it may differ from the accuracy verified on the skin surface. If the region of interest is not accessible, verify in areas as close as possible to the region of interest. 4. In order to avoid nuisance with Spine & Trauma 3D or Orthopedic Navigation software in combination with an affected system camera installed: Do not attempt to use the optional Disposable Clip-On Remote Control. For patient registration, acquire registration points with the standard instrument pivoting method; or for Spine & Trauma 3D use Automatic Image Registration (AIR) if available.

Details

Recall number
Z-2793-2020
Initiated
July 16, 2020
Terminated
July 27, 2021
Product code
OLO – Orthopedic Stereotaxic Instrument
Quantity
9 unit
Distribution
USA: CA, TX, RI OUS: Armenia; Australia; Brazil; Bulgaria; Canada; Finland; France; Germany; Hong Kong; India; Israel; Italy; Japan; Mongolia; Morocco; Spain
Recalling firm
Brainlab AG, Munich

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2792-2020 Class IIBRAINLAB Curve; Model/catalogue (article) numbers for Curve: 19901; 19901B; 19900; 19905…Incorrect manufacturer calibration 07/16/2020Terminated