Recall Z-2803-2026
Argon Medical Devices, Inc · initiated June 15, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
L-Cath PICC S/L Peripherally Inserted Central Catheter: 26ga (1.9F) 0.60mm x 30cm, REF: 384539; 18ga (3.5F) 1.15mm x 60cm, REF: 384464; Basic Kit 28GA (1.2F) x 25cm 1 Lumen), REF: 384516; 20ga (3F) 1.00mm x 60cm, REF: 384463
Codes / lots: REF/UDI-DI/Lot(Expiration): 384539/00886333209934/11488167(7/14/2026), 11488168(7/17/2026), 11494535E(8/1/2026), 11507766(11/13/2026), 11550780(12/12/2026), 11551487(12/15/2026), 11551488(12/15/2026); 384464/00886333209835/11490593(7/28/2026), 11494886(8/28/2026)11501150(10/6/2026); 384516/00886333209910/11500410(8/14/2026), 11501856(8/14/2026), 11504222(9/9/2026), 11506952(9/9/2026), 11523868(9/18/2026), 11551789(9/18/2026), 11553909(11/13/2026), 11559226(11/20/2026); 384463/00886333209828/11490797(8/1/2026), 11490959(8/11/2026)
Reason for recall
Catheters, indicated for short/long-term peripheral access to central venous system for the administration of fluids, medications, nutrients; and sampling, may experience leaking and/or cracking at the hub, which may lead to replacement, inability to flush/aspirate catheter, catheter occlusion, therapy interruption, blood loss, tissue irritation, bloodstream infection.
FDA-determined cause: Device Design
Action taken
On 6/15/2026, recall notices were emailed to customers who were asked to do the following: 1) Provide this recall notice to all people who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2) Segregate, quarantine, and return affected devices to recalling firm. 3) Complete and return the response form via email Arbee.cummings@argonmedical.com For additional information on this letter, please email Scott.Bishop@argonmedical.com
Details
- Recall number
- Z-2803-2026
- Initiated
- June 15, 2026
- Posted by FDA
- July 24, 2026
- Product code
- LJS – Catheter, Intravascular, Therapeutic, Long-Term Greater Than 30 Days
- Quantity
- 6928
- Distribution
- US: AR, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, MA, MD, MI, MN, MO, MS, NC, ND, NJ, NV, NY, OR, TN, TX, UT, VA, WA. OUS: MYS, CAN, SAU
- Recalling firm
- Argon Medical Devices, Inc, Athens, TX
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2804-2026 Class II | L-Cath Midline S/L Catheter 20ga (3F) 1mm x 20cm, REF: 384620Catheters, indicated for short/long-term peripheral access to central venous system for the administration… | 06/15/2026Open, Classified |