Recall Z-2804-2016

Ortho-Clinical Diagnostics · initiated April 19, 2016

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

The VITROS Chemistry Products Calibrator Kit 31, Lot 3155 used in conjunction with: VITROS Chemistry Products HbA1c Reagent Kit, Generation 6 (GEN 6) and Assay Data Disk (ADD), Data Release Versions (DRV) 5873, 5874, 5875, or 5876, is used to calibrate the VITROS 5,1 FS Chemistry System, VITROS 4600 Chemistry System and the VITROS 5600 Integrated System for the determination of percent glycated hemoglobin (HbA1c) in human whole blood.

Codes / lots: Lot 3155 Product Codes: 6842906 & 6801876

Reason for recall

There is an incorrect value (data/calibration mathematics) on ADDs. This incorrect value will prevent a successful calibration of the assay.

FDA-determined cause: Employee error

Action taken

On 19 April 2016, Ortho Clinical Diagnostics distributed Correction Notification letters (Ref. CL2016-087) and ADD DRV 5877 notices to their US consignees via courier service and to their foreign affiliates via email. Customers were advised to install the Assay Data Diskette DRV 5877 or above on their VITROS system. Following a successful calibration and quality control assessment, it is acceptable to use VITROS HbA1c Reagent Kit, GEN 06. Customers who further distributed the product should also provide a copy of the enclosed letter to all customers who received the product. It is suggested that customers complete and return the Confirmation of Receipt form to indicate that you have been informed of this issue. Customers with questions can contact the Technical Solutions Center at 1-800-421-3311.

Details

Recall number
Z-2804-2016
Initiated
April 19, 2016
Terminated
June 8, 2018
Product code
JIT – Calibrator, Secondary
Quantity
520 Units
Distribution
Worldwide Distribution - US Nationwide in the states of: AL, AR, AZ, CA, CT, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, MS, NC, NY, OH, OK, OR, PA, SC, TN, TX, WA, WV & Puerto Rico;** Foreign: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, Republic of Panama, Singapore, Venezuela, England, France, Germany, Italy & Spain.
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY

Official FDA recall record