Recall Z-2834-2016

Alcon Research, Ltd. · initiated July 28, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

LAUREATE World Phaco System

Codes / lots: Lot Numbers: 1658863H, 1805033H, 1805034H, 1817516H, 1817517H, 1826141H, 1826142H, 1857622H, 1857623H, 1868399H, 1868400H

Reason for recall

Steps 16-25 are missing from the LAUREATE Multipak FMS Directions for Use (DFU)

FDA-determined cause: Labeling Change Control

Action taken

All remaining inventory of the identified affected products within Alcon DCs will be supplied corrected directions for use material prior to any further shipments.

Details

Recall number
Z-2834-2016
Initiated
July 28, 2016
Posted by FDA
September 15, 2016
Terminated
June 7, 2017
Product code
HQC – Unit, Phacofragmentation
Quantity
8,646 UNITS
Distribution
Burundi, China, Iran, South Korea, Taiwan, Thailand, USA. *** The recalled product is not sold in the USA or its territories. However, donated units are in use by a reduced number of USA clients for Medical Missions outside the country.
Recalling firm
Alcon Research, Ltd., Fort Worth, TX

Official FDA recall record