Recall Z-2854-2016

EKOS Corporation · initiated August 9, 2016

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

EkoSonic Endovascular Devices EKOS Intelligent Drug Delivery Catheter

Codes / lots: Table 1: Device Identification Information Catalog Number; Lot Number; Serial Number; Expiration Date 500-5518; 160711017; 160711017-005; 06-30-2019 500-5518; 160711017; 160711017-006; 06-30-2019. 500-5518; 160711017; 160711017-003; 06-30-2019. 500-5518; 160711017; 160711017-004; 06-30-2019. 500-5512; 160711042; 160711042-010; 06-30-2019. 500-5518; 160711017; 160711017-001; 06-30-2019. 500-5518; 160711017; 160711017-002; 06-30-2019. 500-5518; 160711017; 160711017-007; 06-30-2019. 500-5518; 160711017; 160711017-008; 06-30-2019. 500-5518; 160711017; 160711017-009; 06-30-2019. 3001627457 08/09/2016-002R CONFIDENTIAL 3 ¿ Name, address, and type of responsible firm on label; number and description of private labels. Manufacturer: EKOS Corporation 11911 North Creek Parkway South Bothell, WA 98011 ¿ Collect a complete copy of all labeling (include product inserts, information sheets.) Applicable labeling is included in Attachment 1. B. CODE: ALL lot and/or serial numbers, batch numbers, catalog numbers, product numbers, packer or manufacturer numbers, etc., exactly as they appear on the product, container, or labeling. There are 10 total devices in distribution. Identification…

Reason for recall

The recalling firm identified ten devices which were distributed and could potentially be non-compliant. These devices potentially underwent a routine manufacturing test with non-compliant parameters, which may result in a small blister in the epoxy over transducers in the treatment zone. It is not known at the time the recall was initiated whether the non-compliant test parameters, or presence of the blisters, adversely affect operation of the device.

FDA-determined cause: Storage

Action taken

All affected consignees were contacted by e-mail on August 9, 2016, and asked to sequester implicated devices.

Details

Recall number
Z-2854-2016
Initiated
August 9, 2016
Terminated
October 5, 2016
Product code
KRA – Catheter, Continuous Flush
Quantity
10 units
Distribution
Three international consignees - Germany, New Zealand and Taiwan
Recalling firm
EKOS Corporation, Bothell, WA

Official FDA recall record