Recall Z-2889-2020

Becton Dickinson & Company · initiated July 28, 2020

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

BD SafetyGlideTM 3ml 25G x 1 3mL SafetyGlide syringes, 50 units in pack

Codes / lots: Catalog #: 305924 Lot #: 8289599 UDI #: 30382903059240

Reason for recall

Printed scale markings on the 3mL SafetyGlide syringes, with catalog # 305924, Lot# 8289599, are printed incorrectly. The scale is skewed to different degrees, resulting in missing and/or partially printed scale numbers and scale lines.

FDA-determined cause: Equipment maintenance

Action taken

BD Sent a customer letter dated 28 July 2020. this letter requests customers to review their inventory for effected lots and Destroy all product subject to the recall. The letter requests customers to return the included response form and notify BD of any adverse events associated to this product.

Details

Recall number
Z-2889-2020
Initiated
July 28, 2020
Terminated
April 23, 2021
Product code
FMI – Needle, Hypodermic, Single Lumen
Quantity
125,200 units
Distribution
US and Canada
Recalling firm
Becton Dickinson & Company, Franklin Lakes, NJ

Official FDA recall record