Recall Z-2912-2016

Siemens Healthcare Diagnostics Inc · initiated September 2, 2016

Class III Terminated

Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws

Product

Siemens RAPIDPoint 400 Measurement Cartridge Siemens SMN numbers:10329756, 10341161

Codes / lots: Serial Numbers:2610900214 through 2621001325 Expiry Dates: Sept 13, 2016 through Dec. 6, 2016

Reason for recall

Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405 Measurement Cartridge

FDA-determined cause: Under Investigation by firm

Action taken

Siemens Healthcare issued An Urgent Field Safety Notice via e-mail to affected Siemens Healthcare Diagnostics Regional offices on September 2, 2016 both in the United States and outside of the United States, for communication with affected customers e: the bias of the chloride values. US accounts sent hard copy via Federal Express and requested to: Complete and return the Field Correction Effectiveness Check form attached to this letter within 7 days. Review this letter with your Medical Director. Questions:Contact your Siemens Customer Care Center or your local Siemens technical support representative.

Details

Recall number
Z-2912-2016
Initiated
September 2, 2016
Posted by FDA
September 30, 2016
Terminated
August 7, 2017
Product code
JFS – Coulometric, Chloride
Quantity
689 cartridges
Distribution
Nationwide: Foreign: Canada, Bahamas,Germany, Japan, Mexico.Belgium, Italy,Spain, Korea, France, Austria, Korea, France, Poland, Australia, china, AE, Bangladesh, Columbia, Czechkoslovakia, Bosnia
Recalling firm
Siemens Healthcare Diagnostics Inc, Norwood, MA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2913-2016 Class IIISiemens RAPIDPoint 405 Measurement Cartridge Siemens SMN numbers: 10844812, 1028322…Error in the value assignment for the Chloride calibrators in the RAPIDPoint 400 and RAPIDPoint 405… 09/02/2016Terminated