Recall Z-2926-2018

Stanbio Laboratory, LP · initiated February 8, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Stanbio AC power adapter for the HemoPoint H2 meter, Model G3000 series. The firm name on the label for the HemoPoint H2 shows Manufactured by EKF-diagnostic GmbH, Germany.

Codes / lots: Serial number range of affected power adapters: 41800001-41804000

Reason for recall

The US prong (type A) adapter plate may crack, break, or detach and remain in the electrical outlet exposing live electrical contacts.

FDA-determined cause: Nonconforming Material/Component

Action taken

The recall was initiated via letter dated 2/7/2018 by first class mail letter on 2/8/2018 instructing the consignee to stop using the adapter if it is affected and it will be replaced.

Details

Recall number
Z-2926-2018
Initiated
February 8, 2018
Terminated
September 4, 2020
Product code
GKR – System, Hemoglobin, Automated
Quantity
2,804 devices
Distribution
Distribution was nationwide. There was military distribution and no government distribution. Foreign distribution was made to Bolivia, Malaysia, Mexico, Papua New Guinea, Paraguay, Philippines, and Suriname.
Recalling firm
Stanbio Laboratory, LP, Boerne, TX

Official FDA recall record