Recall Z-2934-2018

Medical Components, Inc dba MedComp · initiated July 12, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Duo-Flow Double Lumen Soft Tip CRRT/ABP Catheter Set; DL11/24 . Hemodialysis catheter insertion kits

Codes / lots: MCAL840 MCBB800 MCBW450 MCCH690

Reason for recall

The suture wings have the potential to crack or break. Potential patient injury includes catheter dislocation, bleeding, and air emboli. The firm has received six complaints (29 broken suture wings) to date for malfunction, without serious injury or death.

FDA-determined cause: Process control

Action taken

On July 12, 2018, the firm sent a Product Recall letter to its customers. Customers were provided with a full list of product codes and lot numbers and informed that those items had been recalled due to a potential for the suture wings to crack or break. Customers were informed that implanted catheters with intact suture wings do not need to be removed. However, an alternative method of catheter securement should be initiated for all implanted catheters, due to the potential for the suture wings to crack or break. Customers were instructed to return all unused recalled product to Medcomp. If product was further distributed, customers were asked to notify their sub-accounts. Contact your customer service representative for a Returned Goods Authorization (RGA) number if necessary. The phone number to call is 215-256-4201. Customers are asked to complete and return a customer response form included with the recall letter by fax (215-256-9191) or email (complaints@medcompnet.com).

Details

Recall number
Z-2934-2018
Initiated
July 12, 2018
Terminated
July 2, 2020
Product code
MPB – Catheter, Hemodialysis, Non-Implanted
Distribution
Nationwide distribution; worldwide distribution.
Recalling firm
Medical Components, Inc dba MedComp, Harleysville, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2932-2018 Class IIDuo-Flow XTP Straight Full Set; Hemodialysis catheter insertion kits, AAC02106 AAC02107…The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2933-2018 Class IIDouble Lumen CRRT/ABP Catheter Set; Hemodialysis catheter insertion kits, DL 11/15 DL…The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2935-2018 Class IIDuo-Flow IJ Full Tray; MCDLT114IJ MCDLT116IJ MCDLT118IJThe suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2936-2018 Class IIDuo-Flow Catheter Full Tray; MCDLT3500 MCDLT4000 MCDLT4400 MCDLT4424# Hemodialysis…The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2937-2018 Class IIDuo-Flow XTP Straight Full Set, RMS02105 RMS02108The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2938-2018 Class IIDuo-Flow Double Lumen Catheter Only/Catheter Set; T114C T114M T116C T116M T116ME T118C…The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2939-2018 Class IISOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER TRAY; T3500 T4000 T4400 T4424The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2940-2018 Class IISOFT-LINE DUO-FLOW DOUBLE LUMEN CATHETER/CATHETER SET; T94C T94M T96M T98MThe suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2941-2018 Class IIDUO FLOW CUSTOM TRAYS; TRAY #414 TRAY #415 TRAY #67The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated
Z-2942-2018 Class IIDUO FLOW DOUBLE LUMEN CATHETER, XTP114CT, XTP114MT, ; RAULERSON DUO FLOW IJ DOUBLE LUMEN…The suture wings have the potential to crack or break. Potential patient injury includes catheter… 07/12/2018Terminated