Recall Z-2942-2020
MicroPort Orthopedics Inc. · initiated July 31, 2020
Class I – dangerous or defective products that could cause serious health problems or death
Product
PROFEMUR Neck Extra Long Titanium, Catalog Numbers: PHA01206, PHA01236, PHA01256. Hip prosthesis component
Codes / lots: All lots/serial numbers
Reason for recall
MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long Titanium modular femoral neck component after implantation. Worldwide there were 680 Long and X-Long PROFEMUR Titanium modular neck fractures among 123,840 global sales representing is a cumulative rate of modular neck device fracture of 0.55% from 2002 to 2019. Within the United States there were 349 Long and X-Long PROFEMUR Titanium modular neck fractures among 15,786 United States sales representing a cumulative rate of modular neck device fracture with a fracture rate in the United States of 2.21% from 2002 to 2019. There is an average time of 5.4 years to device fracture after implantation.
FDA-determined cause: Device Design
Action taken
The firm initiated the recall by letter on 07/31/2020. The letter provide the following instructions: "Distributors: 1. Please immediately cease use and distribution of these products. 2. Please check your stock, including all consignment stock to immediately determine if you have any affected product. 3. Please complete the form on the last page of this notification and return the form and any affected products to MicroPort Orthopedics per the Actions section of this notification. 4. Please assist us by distributing this notification to all hospitals and surgeons in your territory and collecting and returning completed forms to MicroPort Orthopedics. NOTE: This notice will also be sent directly to the hospitals and surgeons, however they only need to complete the information once, either provided through the distributor or send directly back to MicroPort Orthopedics. Hospitals: 1. Please immediately cease use of these products. 2. Please complete the form on the last page of this notification and return the form and any affected products to MicroPort Orthopedics per the Actions section of this notification. Surgeons: 1. Please immediately cease use of these products. 2. You can continue to monitor your patients per your normal protocol. There are no diagnostic techniques which can predict an unexpected postoperative failure. 3. Should a patient present with sudden pain, instability and difficulty walking/performing common tasks a postoperative fracture can be confirmed via x-ray. 4. Please sign and return this form per the instructions to confirm your notification of this action. ACTIONS: 1. Please complete and return the enclosed form immediately even if you do not have any affected product to return. This information is essential in order to carry out effectiveness checks required per FDA guidelines. Return the completed form by fax to: 901-867-4771, or by e-mail to: PostMarket@ortho.microport.com. Please feel free to
Details
- Recall number
- Z-2942-2020
- Initiated
- July 31, 2020
- Posted by FDA
- September 18, 2020
- Terminated
- April 16, 2024
- Product code
- LWJ – Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented
- Quantity
- 556 units
- Distribution
- worldwide, except China and Japan
- Recalling firm
- MicroPort Orthopedics Inc., Arlington, TN
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2941-2020 Class I | PROFEMUR Neck Long Titanium, Catalog Numbers: PHA01204, PHA01214, PHA01224, PHA01234…MicroPort Orthopedics Inc. states that there have been reports of fractures of the long and extra-long… | 07/31/2020Terminated |