Recall Z-2953-2011
International Technidyne Corp. · initiated May 10, 2011
Product
Hemochron JR. Microcoagulation Systems Prothrombin Time (PT). International Ethidine Corp (ITC)-Nexus Dx. A unitized mlcrocoagulation test intended for use in performing one-stage prothrombin time.
Codes / lots: Catalogue/Code J201 Lot Number: FOJPT059, GOJPT067, GOJPT068, GOJPT069, GOJPT070, GOJPT072, GOJPT073, GOJPT074, GOJPT075, HOJPT076, HOJPT077, HOJPT078, HOJPT079, HOJPT080, HOJPT081, HOJPT082, HOJPT083, JOJPT084, JOJPT085, JOJPT086, JOJPT087, JOJPT088, JOJPT090, KOJPT091, KOJPT092, KOJPT093, MOJPT094, MOJPT095, MOJPT096, MOJPT097, MOJPT098, MOJPT099, A1JPT001, A1JPT002 A1JPT003 A1JPT005 A1JPT007 A1JPT008, A1JPT009, A1JPT010, A1JPT011, A1JPT012, A1JPT013, A1JPT014, A1JPT015, A1JPT016, A1JPT017, A1JPT018, A1JPT019, A1JPT021, A1JPT023, A1JPT024, A1JPT025, B1JPT032, B1JPT033, C1JPT035, C1JPT038, C1JPT041, and C1JPT045.
Reason for recall
Certain lots of Hemochron Jr. PT Cuvettes (J201) demonstrate a higher bias than historically observed when compared to laboratory reference instruments.
FDA-determined cause: Device Design
Action taken
International Technidyne Corp. (ITC) sent "Urgent Medical Device Recall" letters dated May 10, May 25 and June 09, 2011 to all affected customers. The letter included a description of the problem, asked customers fill out the attached form, and return affected product for replacement. For questions, please contact ITC Technical Support at (800) 631-5945.
Details
- Recall number
- Z-2953-2011
- Initiated
- May 10, 2011
- Posted by FDA
- August 5, 2011
- Terminated
- November 2, 2012
- Product code
- JPA – System, Multipurpose For In Vitro Coagulation Studies
- Quantity
- 6,218 boxes in US ; 4713 boxes outside US
- Distribution
- Worldwide Distribution: (USA) Nationwide Distribution including the countries of; Austria, Australia, Canada, Switzerland, Germany, Denmark, Estonia, Spain, Finland, France, United Kingdom, Hong Kong, Ireland, Israel, India, Italy, Korea, Liechtenstien, Libya, Netherlands, Norway, Oman,Portugal, Qatar, Saudi Arabia, Sweden, Turkey, Taiwan, and South Africa.
- Recalling firm
- International Technidyne Corp., Edison, NJ
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2954-2011 | IL GEM PCL/PCL Plus - Gem Test PT Cartridge ITC-Nexus Dx A unitized mlcrocoagulation…Certain lots of Hemochron Jr. PT Cuvettes (J201) and GEM Test PT cartridges demonstrate a higher bias than… | 05/10/2011Terminated |