Recall Z-2955-2026
Philips North America Llc · initiated July 1, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Cardiac Workstation 5000. Product Number: 860439.
Codes / lots: Product Number: 860439. UDI: (01)00884838094826. Serial Numbers: US42510185, US42510184, US52510209, US52510200, US52510226, US52510228, US62510247, US62510250, US62510256, US72510286, US82510299, US82510307, US82510300, US82510313, US82510318, US82510314, US82510335, US82510336, US82510341, US82510325, US92510351, US92510353, US92510354, US92510361, US92510362, US92510370, USN2510413, USN2510414, USN2510434, USN2510435, USN2510437, USD2510506, USD2510519, US12610582, USD2510513, US12610598, US22610609, US22610683, US22610691, US22610689, US32610737, US32610744, US32610759, US32610766, US32610749, US32610739, US32610751, USD2510516, USD2510510, USD2510517, USD2510529, USD2510509, US32610752, US32610753, US32610879, US32610880, US32610901, US32610914, US42610946, US32610915, US32610941, US32610932, US42610943, US42610942, US42610959, US42610958, US42610955, US42610974, US42610967, US42610956, US42610993, US42610960, CN72410032, CN92410061, CNO2410064, CNO2410068, CNN2410075, CND2410080, CND2410101, CND2410103, CN12510116, CN12510136, CN22510140, CN22510142, CN22510146, CN32510148, CN32510161, CN42510168, CN42510181, CN42510183, CN52510204, CN52510201, CN52510217, CN62510239…
Reason for recall
Device will intermittently provide inaccurate data related to native heart rate.
FDA-determined cause: Software design
Action taken
Philips notified consignees on about 07/01/2026 via URGENT Medical Device Correction letter. Consignees were informed that they could continue the use of the device but to refer to the information within the notification letter, disable the automatic Standby feature on the Cardiac Workstation and avoid manually putting the Cardiac Workstation in Standby mode to minimize the inaccurate heart rate value from occurring, and to pass the notice to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Additionally, consignees were requested to place the Urgent Medical Device Correction letter with the documentation of the Cardiac Workstation and associated devices in a place where it is most likely to be seen and viewed. Philips representatives will contact consignees to arrange an upgrade to the Cardiac Workstation Release 1.2.0.7.
Details
- Recall number
- Z-2955-2026
- Initiated
- July 1, 2026
- Posted by FDA
- August 14, 2026
- Product code
- DPS – Electrocardiograph
- Quantity
- 796 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, France, Germany, Gibraltar, Hong Kong, Iceland, Ireland, Iraq, Italy, Luxembourg, Malta, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Korea, Spain…
- Recalling firm
- Philips North America Llc, Cambridge, MA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2956-2026 Class II | Cardiac Workstation 7000. Product Number: 860441.Device will intermittently provide inaccurate data related to native heart rate. | 07/01/2026Open, Classified |