Recall Z-2958-2026

Arrowhead DE LLC · initiated June 23, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Arrow-LOK Implant, REF: 21-253519, 21-253522, 21-253525, 21-253528, 21-253530, 21-253533, 21-253535, 21-253540, 21-253545, 21-253550, 21-353513, 21-353516, 27-253519, 27-253522, 27-253525, 27-253528, 27-253530, 27-253533, 27-253535.

Codes / lots: REF/UDI-DI/Lot: 21-253519/00855967006035/24662119, 242162; 21-253522/00855967006042/24662122, 242163, 220702; 21-253525/00855967006059/24662125, 220703, 242164; 21-253528/00855967006066/24662128, 242165, 230704; 21-253530/00855967006073/242166, 230705, 220705; 21-253533/00855967006080/24662133, 242167, 230706; 21-253535/00855967006097/24662135, 242168, 230707; 21-253540/00855967006110/242169; 21-253545/00855967006127/242170; 21-253550/00855967006134/24662150; 21-353513/00855967006004/24662113, 242160; 21-353516/00855967006011/24662116, 230701, 242161; 27-253519/00855967006028/24662719, 242171; 27-253522/00855967006141/242172; 27-253525/00855967006158/242173; 27-253528/00855967006165/242174; 27-253530/00855967006172/242175; 27-253533/00855967006189/24662733, 242176; 27-253535/00855967006196/24662735, 242177

Reason for recall

Implant indicated for fixation of osteotomies, arthrodeses and reconstruction in the lesser toes following corrective procedures, did not undergo required gamma radiation dose audit. A potential reduction in sterility assurance could lead to patient infection.

FDA-determined cause: Process change control

Action taken

On 6/23/2026 customers were phoned and informed to return affected implants. On 7/7/2026, recall notices were mailed to risk managers who were asked to do the following: 1) Cease use, distribution, and quarantine affected product. 2) Contact firm customer service to obtain instructions for returning product. 3) Email mail tracking numbers to firm. 4) Complete and return the acknowledgement and receipt form. Questions can be directed to the firm at 470-419-1429

Details

Recall number
Z-2958-2026
Initiated
June 23, 2026
Posted by FDA
August 14, 2026
Product code
HTY – Pin, Fixation, Smooth
Quantity
4,171
Distribution
US Nationwide distribution in the states of VT, FL, TN, NY, GA, TX, HI.
Recalling firm
Arrowhead DE LLC, Peachtree City, GA

Official FDA recall record