Recall Z-2968-2026

Gentherm Medical, LLC · initiated July 13, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Blanketrol II Model 222S; Thermal regulating system.

Codes / lots: UDI: 1061303186165 Serial Numbers:131-2-00580 through 262-3-20277

Reason for recall

Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.

FDA-determined cause: Device Design

Action taken

Genthem issued an URGENT Medical Device Notice to its consignees on 7/17/2026 via email. The notice explained the issue, potential risk, and request the following: "Interim Recommendations: 1. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. 2. Users should review and follow the device Instructions for Use; users should routinely verify the device s function and patient s condition during use. 3. Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice." For questions or if you have experienced the described issue or a related adverse event, call Gentherm Medical Technical Support at 1-888-437-5608.

Details

Recall number
Z-2968-2026
Initiated
July 13, 2026
Posted by FDA
August 14, 2026
Product code
DWJ – System, Thermal Regulating
Quantity
182 units
Distribution
Worldwide - US Nationwide distribution including in the territories of Puerto Rico, and the countries of Brazil, Canada, Chile, Colombia, Costa Rica, El Salvador, Guatemala, Mexico, Republic of Panama, Saudi Arabia, Taiwan, Thailand, Bahamas, and Trinidad and Tobago.
Recalling firm
Gentherm Medical, LLC, Blue Ash, OH

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2969-2026 Class IIInnercool Coolblue Console, Model STx; Thermal regulating system.Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and… 07/13/2026Open, Classified
Z-2970-2026 Class IIBlanketrol III, Model 233; Thermal regulating system.Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and… 07/13/2026Open, Classified