Recall Z-2971-2026
Paragon 28, Inc. · initiated July 10, 2026
Class III – products unlikely to cause any adverse health reaction but that violate FDA labeling or manufacturing laws
Product
Custom Made Device Implant 6123 Navicular Cage Small Right Rev A, Model/Catalog Number 6123S R; foot implant
Codes / lots: Lot Number AOCIA6123
Reason for recall
Device was recalled due to an elevated oxygen level in excess of the allowable limit. Elevated oxygen levels may contribute to reduced material ductility.
FDA-determined cause: Under Investigation by firm
Action taken
Paragon 28 issued an URGENT MEDICAL DEVICE RECALL to its consignees on 7/14/2026 via FedEx 2nd day. The notice explained the issue, potential risk, and requested the following: Distributors: "Your Responsibilities 1. Review this notification and ensure that affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Immediately return all affected product from your distributorship and from affected hospitals within your territory. a. Complete Attachment 1 Inventory Return Certification Form for each return and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form must be returned even if you do not have affected products available to return in your territory. b. Include a hardcopy of Attachment 1 in each carton of your return shipment for immediate processing. c. Mark RECALL on the outside of the returned cartons. 4. Retain a copy of your Inventory Return Certification and product return forms for your records in the event of a compliance audit of your facility. 5. If you have further questions or concerns after reviewing this notice, please call 888-728-1888. Alternatively, your questions may be emailed to CorporateQuality.PostMarket@zimmerbiomet.com." Risk Manager: "Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Paragon 28 sales representative and quarantine all affected product. Your Paragon 28 sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with the field safety notice for hospitals and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. This form shall be returned even if you do not have affected products at your facility. 5. Retai
Details
- Recall number
- Z-2971-2026
- Initiated
- July 10, 2026
- Posted by FDA
- August 14, 2026
- Product code
- PLF – Bone Wedge
- Quantity
- 1
- Distribution
- US Nationwide distribution i the state of MT.
- Recalling firm
- Paragon 28, Inc., Englewood, CO
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|