Recall Z-2980-2026

Medtronic MiniMed, Inc. · initiated July 13, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712;

Codes / lots: Model/UDI-DI/Serial Numbers: MMT-1711/00763000404550/NG2660048H, NG2663343H, NG3015306H; MMT-1712/00763000404567/NG2628486H, NG2758213H, NG2999651H, NG3349198H, NG3351719H, NG2678842H, NG2831821H, NG2831546H, NG2932027H, NG3049657H, NG3033094H, NG2831538H, NG3738739H

Reason for recall

Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump case or keypad, allow moisture ingress, and result in insulin over- or under-delivery, potentially causing hypoglycemia, hyperglycemia, or diabetic ketoacidosis.

FDA-determined cause: Process design

Action taken

On 7/13/2026, field safety notices were sent to customers asking them to do the following: 1) Stop using the loaner pump and contact the firm to arrange for loaner pump return. 2) Contact your Healthcare Professional (HCP) to ensure you have access to backup insulin therapy supplies or alternative plans, which may include short- and long-acting insulins, syringes, with dosage instructions from your HCP. 3) Complete and return the customer acknowledgement form. Questions can be directed to the firm's Tech Support: 1-800-646-4633

Details

Recall number
Z-2980-2026
Initiated
July 13, 2026
Posted by FDA
August 18, 2026
Product code
QFG – Alternate Controller Enabled Insulin Infusion Pump
Quantity
16
Distribution
International distribution to the countries of Germany, Denmark, Austria, Netherlands, United Kingdom, Italy, Belgium, Ireland, Switzerland, Sweden, Northern Ireland.
Recalling firm
Medtronic MiniMed, Inc., Northridge, CA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-2981-2026 Class IIMiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;Loaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump… 07/13/2026Open, Classified
Z-2982-2026 Class IIMiniMed 780G Insulin Pump, Model: MMT-1885CLW, MMT-1886CLWLoaner insulin pumps may have been cleaned with chemically incompatible products that could damage the pump… 07/13/2026Open, Classified