Recall Z-2993-2011
Varian Medical Systems, Inc. Oncology Systems · initiated July 13, 2011
Product
GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 320 mm. Remote controlled radionuclide applicator system. Varian Medical Systems, Inc. Mfg by Varian Medical Systems Haan GmbH, Haan, Germany Product Usage: The Interstitial Plastic Needles with 2 mm diameter are designed for interstitial treatments in areas such as head and neck, gynecology, breast or prostate. In order to avoid bending and to facilitate the introduction of the needles each needle is supplied with a Mandrin
Codes / lots: Part numbers GM11007580; All lots C01 to D01 and lots H and I
Reason for recall
Needle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal.
FDA-determined cause: Device Design
Action taken
Varian sent an "Urgent Medical Device Correction/Urgent Field Safety Notice" letter dated July 13, 2011 to all possibly affected customers. This notification supplements and modifies the previous "Customer Technical Bulletins" originally sent in September 2008. The initial Customer Technical Bulletins informed customers of the possibility that needle tips cans crack and separate. It also reminded users that the product has a limited life and to check needles prior to use as per instructions. Current correction notice letter instruct customers to stop using any plastic needles in inventory with lot numbers C01 to D01 and those with lot numbers starting with H or I. The letter also request that the recalled products be returned to Varian BrachyTherapy for replacement. The letters include information as to where the lot number is located on the device. All appropriate personnel should be notified of the correction notice. Contact information is provided for all geographical areas. USA contact for questions or concerns is 1-800-360-7909.
Details
- Recall number
- Z-2993-2011
- Initiated
- July 13, 2011
- Posted by FDA
- August 10, 2011
- Terminated
- June 20, 2012
- Product code
- JAQ – System, Applicator, Radionuclide, Remote-Controlled
- Quantity
- 601 - all sizes.
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of: Argentina, Austria, Bangladesh, Belgium, Brazil, Canada, Columbia, Cyprus, Czech Republic Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, India, Japan, Kazahastan, Kuwait, Lebanon, Macedonia, Mexico, Moldova, Nepal New Zealand, Norway, Pakistan, Philippines, Poland, Russia, Serbia, Singapore, Slovenia, South…
- Recalling firm
- Varian Medical Systems, Inc. Oncology Systems, Palo Alto, CA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-2991-2011 | GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 113 mm. Remote controlled…Needle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal. | 07/13/2011Terminated |
| Z-2992-2011 | GammaMed Plastic Needle with Mandrin; 2.0mm diameter, length 200 mm. Remote controlled…Needle tip may separate from the needle shaft such that the tip may detach in the patient upon withdrawal. | 07/13/2011Terminated |