Recall Z-3030-2011

Tosoh Bioscience Inc · initiated May 25, 2011

Terminated

Product

Hemoglobin F & A2 Calibrator, 0.2 mL, 4 brown glass vials/box. Mfr. By: Tosoh Bioscience, Minato-ku, Tokyo, Japan. Hemoglobin F & A2 Calibrator is used as in-vitro calibration material for Beta-thalassemia testing on the Tosoh G7.

Codes / lots: Part Number 020019. Lot # BCAL05R. Used By Date 2011-10.

Reason for recall

The value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the label is 5.7%. The criteria of the value is +/- 0.3%.

FDA-determined cause: Process control

Action taken

TOSOH Bioscience sent an "Urgent Recall Notification" letter dated July, 25, 2011 to all affected customers. The letter described the problem and asks customers to immediately discontinue use of the product and return all remaining product back to the firm. Customers can contact TOSOH for replacement. For additional information or question on this recall please contact TOSOH Technical Support at (800) 248-6769.

Details

Recall number
Z-3030-2011
Initiated
May 25, 2011
Posted by FDA
August 17, 2011
Terminated
February 24, 2012
Product code
JPD – Hemoglobin A2 Quantitation
Quantity
6 boxes/4/2 mL btls
Distribution
Worldwide Distribution: Nationwide (USA) distribution including the states of FL and IL; and the country of Canada.
Recalling firm
Tosoh Bioscience Inc, Grove City, OH

Official FDA recall record