Recall Z-3050-2017

Alcon Research, Ltd. · initiated June 30, 2017

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

ORA System with VerifEye, Catalog Number 8065998300 For use during intraocular lens surgery

Codes / lots: Serial Numbers: C0002,C2006,C2020,C2026,C2054,C3009R,C3010R,C3024,C3036,C3066,C3067,C3085R,C3090,C3093,C3108,C3136,C3153,C4001,C4003,C4005,C4006,C4007,C4009,C4010,C4011,C4012,C4013,C4014,C4015,C4016,C4017,C4019,C4020,C4021,C4022,C4023,C4024,C4025,C4026,C4027,C4028,C4029,C4030,C4031,C4035,C4036,C4037,C4038,C4039,C4040,C4043,C4045,C4046R,C4047R,C4049R,C4050R,C4052R,C4053R,C4054R,C4055R,C4056R,C4057R,C4058R,C4059R,C4060R,C4061R,C4062R,C4065R,C4066R,C4067R,C4068R,C4069R,C4070R,C4071R,C4072R,C4073R,C4074R,C4075R,C4076R,C4077R,C4078R,C4079R,C4080R,C4083R,C4085R,C4086R,C4087RC4088R,C4089R,C4090R,C4091R,C4092R,C4093R,C4094R,C4095R,C4096R,C4097R,C4098R,C4099R,C4100R,C4101R,C4102R,C4103R,C4104R,C4105R,C4106R,C4107R,C4108R,C4109R,C4110R,C4111R,C4112R,C4113R,C4115R,C4117R,C4118R,C4119R,C4120R,C4121R,C4122R,C4123R,C4124R,C4126R,C4127R,C4128R,C4129R,C4130R,C4132R,C4133R,C4134R,C4135R,C4136R,C4138R,C4140,C4142,C4143,C4144,C4145,C4146,C4147,C4148,C4149,C4150,C4151,C4152R,C4153,C4154,C4155,C4156,C4157R,C4158,C4159,C4160R,C4161R,C4162R,C4163R,C4164R,C4165R,C4166R,C4167R,C4168R,C4169R,C4170R,C4173R,C4174R,C4175,C4177R,C4178R,C4179R,C4180R,C4181R,C4182R,C4183R,C4184R,C4185R,C4186R,C4187R,C4188R,C4189…

Reason for recall

Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This issue appears to have been caused by a software coding error that results in the lens coefficients for an IOL model being downloaded from the Alcon server in an incorrect order.

FDA-determined cause: Software design

Action taken

The Market Action was initiated with Initial telephone contact to all affected customers starting June 30, 2017. These customers were informed of the issue, and advised not to use their ORA Carts for calculations with the lens model affected on their specific system. A confirmatory letter (Attachment 4) was also sent via overnight mail to these customers July 12, 2017. As of July 12, 2017, Alcon has identified 8 ORA Carts that are affected by the software coding error. Alcon has reset the IOL databases on all 8 of the identified affected ORA Carts.

Details

Recall number
Z-3050-2017
Initiated
June 30, 2017
Terminated
August 27, 2019
Product code
NCF – Aberrometer, Ophthalmic
Quantity
367 units
Distribution
Worldwide distribution. US Nationwide and Argentina, Australia, Belgium, Brazil, Canada, China, Colombia, France, India, Japan, Mexico, Netherlands, Panama, Portugal, Romania, Spain, Thailand, UAE, and United Kingdom
Recalling firm
Alcon Research, Ltd., Fort Worth, TX

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3049-2017 Class IIORA System with VerifEye+ Cart, Catalog Number 8065998307 For use during intraocular…Some ORA Carts have the potential to return an incorrect IOL power measurement during cataract surgery. This… 06/30/2017Terminated