Recall Z-3050-2026

Medline Industries, LP · initiated June 12, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Medline Convenience Kits: 1) DYKM1414, Catalog Number: DYKM1414; 2) DYKM1414, Catalog Number: DYKM1414; 3) DYNDH1462A, Catalog Number: DYNDH1462A; 4) DYNJ38145F, Catalog Number: DYNJ38145F; 5) DYNJ63199A, Catalog Number: DYNJ63199A

Codes / lots: 1) DYKM1414, UDI-DI: 10889942481293(each), 40889942481294(case), Lot Number: 25BBB327; 2) DYKM1414, UDI-DI: 10889942481293(each), 40889942481294(case), Lot Number: 24KBE261; 3) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24ABC301; 4) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24ABT704; 5) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24BBO246; 6) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24BBU663; 7) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 24CBJ141; 8) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23JBH263; 9) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23JBO539; 10) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23KBG333; 11) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23KBV434; 12) DYNDH1462A, UDI-DI: 10193489993424(each), 40193489993425(case), Lot Number: 23LBM123; 13) DYNJ38145F, UDI-DI: 10193489277173(each), 40193489277174(case), Lot Number: 25ABS139; 14) DYNJ38145F, UDI-DI:…

Reason for recall

The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach of sterility in the packaging, due to wrinkles in the paper lidding which may extend to the seal area. If the sterile barrier is compromised, patients and users may be exposed to microbial contamination. Potential health consequences may include no injury, localized infection, or, less commonly, more serious infection requiring medical treatment, antimicrobial therapy, procedural intervention, or hospitalization. In rare cases, particularly in critically ill or immunocompromised patients, severe infection could become life-threatening.

FDA-determined cause: Nonconforming Material/Component

Action taken

An URGENT MEDICAL DEVICE RECALL notification letter dated 6/12/25 was sent to customers. REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-26-097-FGX1 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or recalls@medline.com.

Details

Recall number
Z-3050-2026
Initiated
June 12, 2026
Posted by FDA
August 31, 2026
Product code
OFF – Central Venous Catheter Tray
Quantity
58037 kits
Distribution
US Nationwide distribution.
Recalling firm
Medline Industries, LP, Northfield, IL

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3048-2026 Class IIMedline Convenience Kits: 1) DRAWER 9 CRICOTHYROTOMY KIT, Model Number: ACC010741; 2)…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3049-2026 Class IIMedline Convenience Kits: 1) CDS840396X, Catalog Number: CDS840396X; 2) CDS840396Y…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3051-2026 Class IIMedline Convenience Kits: 1) IVS BLOOD CULTURE KIT, Model Number: DYNDH1920; 2) IVS…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3052-2026 Class IIMedline Convenience Kits: 1) COLORECTAL ME, Model Number: DYNJ906437DThe kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3053-2026 Class IIMedline Convenience Kits: 1) ULTRASOUND BREAST PACK, Model Number: DYNJ63375CThe kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3054-2026 Class IIMedline Convenience Kits: 1) IV START KIT, Catalog Number: DYNDV2520; 2) PERIPHERAL LAB…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3055-2026 Class IIMedline Convenience Kits: 1) ARTERIAL LINE INSERTION KIT, Model Number: ART1295The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3056-2026 Class IIMedline Convenience Kits: 1) IV START KIT LAB DRAW-SH, Model Number: CDS860014P; 2) IV…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3057-2026 Class IIMedline Convenience Kits: 1) BLOOD CULTURE TRAY PEDIATRIC, Model Number: DYNDH1193; 2)…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3058-2026 Class IIMedline Convenience Kits: 1) PRE-OP KIT, Catalog Number: DYKS1043F; 2) PRE OP KIT…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3059-2026 Class IIMedline Convenience Kits: 1) LABOR CDS, Model Number: CDS981762B; 2) VAG DELIVERY…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3060-2026 Class IIMedline Convenience Kits: 1) C-SECTION PREP, Model Number: DYNJ32702BThe kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3061-2026 Class IIMedline Convenience Kits: 1) EYE KIT, Model Number: DYKM2751; 2) EYE KIT, Model Number:…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3062-2026 Class IIMedline Convenience Kits: 1) ORTHO ASPIRATION KIT, Model Number: DYKM1636A; 2) SHOULD…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified
Z-3063-2026 Class IIMedline Convenience Kits: 1) BLOOD DRAW KIT, Catalog Number: DYLAB1081; 2) WHOLE BLOOD…The kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which have a potential breach… 06/12/2026Open, Classified