Recall Z-3088-2026

Medos International Sarl · initiated July 23, 2026

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Velys Enabling Tech Robotic-Assist Station REF: 451611001

Codes / lots: UDI: 10705034645995/Serial Numbers: W14316295301 W15316297501 W15316298001 W17316322901 W17316323301 W21316347401 W21316347501 W21316347601 W21316347801 W21316347901 W21316348001 W22316365801 W34316479001 W34316479101 W34316479201 W34316479301 W34316479401 W34316479601 W34316479801 W34316480001 W34316480101 W49316554301 W51316597001 W51316597101 Y03316603601 Y03316603701 Y03316603801 Y07316625901 Y15316690901

Reason for recall

Software issue within robotic controller which may cause unintended movement of the robotic arm during bone preparation or instrument positioning.

FDA-determined cause: Under Investigation by firm

Action taken

On July 23, 2026, DePuy Synthes contacted affected consigenees via phone call concerning the recall. On July 24, 2026 Depuy issued a Urgent Medical Device Correction notification via sales associate visit. DePuy ask consignees to take the following actions: 1. Do not use the subject VELYS ROBOTIC-ASSIST STATION and quarantine it immediately. The VELYS. 2. Navigation Station and VELYS Camera Station can continue to be used following the Instructions for Use. 3. Complete, sign, and return the attached Business Response Form to OneMD-Field-Actions@its.jnj.com within three (3) business days of receipt of this notification. 4. Forward this notice to anyone in your facility that needs to be informed 5. Post a copy of this notice in a visible area for awareness and keep a copy for your records. 6. As with any medical device, adverse reactions or quality problems experienced with the use of this product should be reported to the FDA's MedWatch Adverse Event Reporting Program Online: https://www.fda.gov/safety/medwatch-fda-safety-information-and-adverse-event-reporting-program/reportingserious

Details

Recall number
Z-3088-2026
Initiated
July 23, 2026
Posted by FDA
September 2, 2026
Product code
OLO – Orthopedic Stereotaxic Instrument
Quantity
29 stations
Distribution
US Nationwide distribution in the states of DE, FL, IL, KY, MD, MI, MN, NJ, NM, NY, OR, PA, TN, TX, VA, WA.
Recalling firm
Medos International Sarl, Le Locle