Recall Z-3114-2026
C.R. Bard Inc · initiated June 25, 2026
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
Products that contain the Pericare Wipes: Bardex Lubricath Foley Catheter Tray REF: 899914 899916 899918 UDI codes: 00801741024801 10801741024808 00801741024818 10801741024815 00801741024825 10801741024822 SureStep" Foley Tray System with Bardex Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep" Foley Tray System Bard Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep" Foley Tray System Bard Lubricath Foley Catheter Tray REF: A899616 UDI code: 00801741104534 10801741104531 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899916 UDI code: 00801741074011 10801741074018 SureStep" Foley Tray System Bardex Lubricath Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978 10801741120975 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902914 A902916 A902918 UDI code: 0080174107406610801741074063 00801741074073 10801741074070 0080174107408010801741074087 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 304916 UDI code: 00801741019736 01801741019739 Bard Urine Meter Foley Catheter Tray With Bardex Lubricath Foley Catheter REF: 902916 UDI codes: 00801741024863 10801741024860 (416A) SureStep Foley Tray System Bardex IC Complete Care (714A/716A/718A)SureStep Foley Tray System Bardex IC Complete REF: A304914 A304916 A304918 UDI codes: 00801741104350 10801741104357 00801741104367 10801741104364 00801741100499 10801741100496 SureStep" Foley Tray System Bard Lubricath Foley REF: A800365 UDI codes: 00801741073960 10801741073967 SureStep" Foley Tray System Bardex Lubricath Foley Catheter REF: A899918 UDI codes: 00801741074028 10801741074025 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902618 UDI codes: 00801741104640 10801741104647 SureStep" Foley Tray System BardLubricath Foley Catheter REF: A902616 UDI codes: 00801741104633 10801741104630 SureStep Intermittent Catheter Tray REF: INTL16C UDI codes: 00801741183195 10801741183192 For urological use only, Intended for use in the drainage and/or collection and/or measurement of urine.
Codes / lots: Catalog Numbers / UDI codes: 899914 / 00801741024801¿ 10801741024808 Lot Number: NGKU4611 899916 / 00801741024818¿ 10801741024815 Lot Numbers: NGKU2448 NGKU4610 899918 / 00801741024825¿ 10801741024822 Lot Number: NGKU3555 ***** SureStep" Foley Tray System with Bardex¿ Lubricath Temp Sensing Foley Catheter REF: A119416M UDI code: 00801741073700 10801741073707 SureStep" Foley Tray System Bard¿ Lubricath Foley REF: A800361 UDI code: 00801741073939 10801741073936 SureStep" Foley Tray System Bard¿ Lubricath¿ Foley Catheter Tray REF: A899616 UDI code: 00801741104534¿ 10801741104531 SureStep" Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899914 UDI code: 00801741074004 10801741074001 SureStep" Foley Tray System Bardex¿ Lubricath¿ Foley Catheter REF: A899916 UDI code: 00801741074011¿ 10801741074018 SureStep" Foley Tray System Bardex¿ Lubricath¿ Temperature Sensing Catheter REF: A900216 UDI code: 00801741120978¿ 10801741120975 SureStep" Foley Tray System Bard¿Lubricath¿ Foley Catheter REF: A902914 A902916 A902918 UDI code: 00801741074066¿10801741074063 00801741074073 10801741074070 00801741074080¿10801741074087 Bard¿ Urine Meter Foley Catheter Tray With Bardex¿ Lubricath¿…
Reason for recall
Due to complaints received indicating presence of contamination on the surface of pericare wipes provided with various Foley and Intermittent Catheter Trays.
FDA-determined cause: Under Investigation by firm
Action taken
On 06/25/2026, an "URGENT Medical Device Recall" Letter was via regional courier services or other electronic means to inform customers that based on customer feedback, BD has identified that the CliniClean Pericare Wipes provided within Urinary Catheter Trays may be potentially contaminated with an unknown non-biological substance. Customer are instructed to: Do not use CliniClean Pericare Wipes contained within the affected products listed in Attachment 2. The CliniClean Pericare Wipes are the only impacted component within the tray and all remaining components can continue to be safely used. Users should source alternative, cleansing wipes for the impacted product lots listed in Attachment 2 and ensure they continue to follow their facility protocols for providing perineal care for patients with urinary catheters. If affected CliniClean Pericare Wipes have been used without issue, no additional treatment or clinical follow up is required. 1. For affected lots, remove and destroy wipes within the product package at the point of use and do not use. Source alternative cleansing wipes for all impacted urinary catheter trays listed in Attachment 2. 2. Share this notice with any users of the product within your facilities to ensure they are also aware of this Medical Device Correction. 3. Post this notice in all locations where inventory of these trays are stored. 4. Complete and return the attached Customer Response Form to the BD contact noted on the form confirming acknowledgement of the correction notification, whether or not you have any affected product, so that BD may confirm your receipt of this notification. For Questions - contact: BD Medical Information Services: Available 24/7 Select #2 when prompted 1-800-555-7422 or Email: medical.information@bd.com BD Technical Support 1-844-8BD-LIFE (1-844-823-5433) Say Product Complaint when prompted M-F 8am - 5pm CT or Email: productcomplaints@bd.com
Details
- Recall number
- Z-3114-2026
- Initiated
- June 25, 2026
- Posted by FDA
- September 4, 2026
- Product code
- EZC – Catheter, Coude
- Quantity
- 12,463 trays/kits
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI and WY. The countries of Bahrain, Canada, Costs Rica, Guatemala, Mexico, Saudi Arabia, and United Arab Emirates.
- Recalling firm
- C.R. Bard Inc, Covington, GA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-3115-2026 Class II | Products that contain the Pericare Wipes: Bard Foley Tray With Lubri-Sil All-Silicone…Due to complaints received indicating presence of contamination on the surface of pericare wipes provided… | 06/25/2026Open, Classified |
| Z-3116-2026 Class II | Products that contain the Pericare Wipes: *Bardex I.C. Complete Care Foley Catheter Tray…Due to complaints received indicating presence of contamination on the surface of pericare wipes provided… | 06/25/2026Open, Classified |
| Z-3117-2026 Class II | Products that contain the Pericare Wipes: SureStep Intermittent Catheter Tray REF:…Due to complaints received indicating presence of contamination on the surface of pericare wipes provided… | 06/25/2026Open, Classified |
| Z-3118-2026 Class II | Products that contain the Pericare Wipes: SureStep" Foley Tray System Add-A-Foley…Due to complaints received indicating presence of contamination on the surface of pericare wipes provided… | 06/25/2026Open, Classified |
| Z-3119-2026 Class II | Products that contain the Pericare Wipes: (Canada Only) SureStep" Foley Tray System…Due to complaints received indicating presence of contamination on the surface of pericare wipes provided… | 06/25/2026Open, Classified |