Recall Z-3171-2026
GE Healthcare · initiated July 31, 2026
Product
Core Services Hub V8, a set of backend services including Cross Enterprise (XE) and Workflow Core (WFC) services that enables enterprise imaging workflows, All versions
Codes / lots: System IDs: CARTI27326018 KGM24849767P2 KGM26790380P1 KGM26790386T1 KGM26790388T2 PIPELINEH26974665 PIPELINEH27036913 KRM25208408 26885371TenantCSH 26897935StgTenCSH VVMC25199249 Not Available VVMC26130320 DDOC26923416 BAYHEALTH26580398 BAYHEALTH26871765 JFKR27127669 LAKESIDEM27124493 BYH25222926 LMC427127671 KRMC27127670 MH826900861 ICA27062095 NORTHWEST24866978-XE NORTHWEST27294841-XE NORTHWEST27294843-XE JHSJHOCC24849766 ANN & ROBE24866975 JHSJHOCC27005469 GOSH26799709CSHP GOSH26848483CSHT CMAE26304473 CMAE26774289 MSUHEALTH26980323 Not Available CHILDRENS24808132 XE VCP 103829-6-CoreServices Hub-XE VMP 103829-6-CoreServices Hub-XE Test 103829-6-CoreServices Hub-WFC VCP 103829-6-CoreServices Hub-WFC VMP 103829-6-CoreServices Hub-WFC Test OMCR25206023 ETP Prod CoreServices 1 ETP Prod CSH 2 CHILDRENS25171538 26885369TenantCSH BC325168126 BC324829461 BC325239077 BC325239084 BC326538517 NCDC26959094 MINNETOHE26945472 AMIM27062090 VTH26323581 SHOREMEMO24810482 SHOREMEMO24988230 SHOREMEMO25249340 VT27062091 NNAR25264207 107500-1-CoreServices Hub-27003899 UNIVERSITY26715185 MSI25263844 RRA24845093 RICHMONDU26628093 MFCI24810480 MFCI24825168 LMH25269182 TIET24966115 UTMC25263843…
Reason for recall
Under certain conditions historical imaging data migrated from a legacy system may become incorrectly associated with a different patient. If this occurs, studies or images may be displayed under the wrong patient record, which could lead to misdiagnosis or a delay in patient care.
FDA-determined cause: Under Investigation by firm
Action taken
An URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/31/26 was sent to customers. Actions to be taken by Customer/User Pending corrections from GE HealthCare, you can continue to use your systems by following the instructions below. 1. Do not perform any new inbound data (DICOM) migration projects. Please contact GE HealthCare before starting any new migration projects. 2. If using Enterprise Archive with Core Services Hub/Workflow Core Services, when reviewing or exporting prior imaging studies after migration to Enterprise Archive, verify that the study and images within the study are associated with the correct patient as follows: At the Study level: " Compare name, date of birth, and patient identifiers only if burned-in or embedded in the image. Do not rely on patient information in the image overlays that can be toggled on and off. " If burned-in or embedded patient identifiers are not available, compare with prior studies based on patient anatomy, implants, or unique anatomical features. At the Image level: " Confirm that the images within the study do not belong to different patients based on acquisition date and time or other patient identifiers. Please maintain any previous source systems or migration servers pending correction of this issue by GE HealthCare. Please ensure that all potential users in your facility are made aware of this safety notification and the recommended actions above. Please retain this document for your records. Please complete and return the attached acknowledgement form electronically via FMI 85492 Digital Response Form or print, fill out manually, scan, and email to recall.85492@gehealthcare.com. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.
Details
- Recall number
- Z-3171-2026
- Initiated
- July 31, 2026
- Posted by FDA
- September 15, 2026
- Product code
- LLZ – System, Image Processing, Radiological
- Quantity
- 250 systems
- Distribution
- Worldwide - US Nationwide distribution.
- Recalling firm
- GE Healthcare, Chicago, IL
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-3166-2026 | Centricity Enterprise Archive V2, V3, and V4, software product for receiving, archiving…Under certain conditions historical imaging data migrated from a legacy system may become incorrectly… | 07/31/2026Open, Classified |
| Z-3167-2026 | Enterprise Archive V8, software product for receiving, archiving, and sending electronic…Under certain conditions historical imaging data migrated from a legacy system may become incorrectly… | 07/31/2026Open, Classified |
| Z-3168-2026 | Centricity PACS V2, V3, V4, V6, V7, software product is intended for the storage…Under certain conditions historical imaging data migrated from a legacy system may become incorrectly… | 07/31/2026Open, Classified |
| Z-3169-2026 | Centricity PACS Server V7, intended for managing digital medical images and associated…Under certain conditions historical imaging data migrated from a legacy system may become incorrectly… | 07/31/2026Open, Classified |
| Z-3170-2026 | Workflow Core Services V8, intended to provide set of services for storage and retrieval…Under certain conditions historical imaging data migrated from a legacy system may become incorrectly… | 07/31/2026Open, Classified |