Recall Z-3187-2018

Tecan US, Inc. · initiated July 23, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Freedom EV0-2 with any version of EVOware and EVOware Sample Tracking Add-On V1.0 or higher

Codes / lots: Part Numbers for Models of Freedom EVO with any version of EVOware: a. Freedom EVO 100: 10641100, 30020010, 30032010; b. Freedom EVO 150: 10641150, 10641152, 30020015, 30032015; c. Freedom EVO 200 Base Unit: 10641200, 10641202, 30020020, 30032020; Part Numbers for Software: EVOware Sample Tracking Add-On for Multi¿ Instrument Configurations: 30019971; Serial Numbers for device: 1302001540, 1708006808, 1208007284, 1503007126, 1503007304, 905000588, 1205002755, 1205002802, 1305007607, 1308005431, 1310005144, 1310005145, 905000587, 1705007973, 904000335, 1510009082, 1109004352, 1704005194, 1007000504, 1305002594, 1307006551 and 1105004439.

Reason for recall

There is a software issue that can potentially impact sample identification and traceability under specific conditions when using EVOware Add On Software: Sample Tracking -Multi Instrument V1.0 or higher, Part Number 30019971.

FDA-determined cause: Device Design

Action taken

On July 23, 2018, Tecan sent Field Corrective Action Letters to their consignees advising them of the problem. Field corrections will be made in the near future.

Details

Recall number
Z-3187-2018
Initiated
July 23, 2018
Terminated
August 26, 2020
Product code
JQW – Station, Pipetting And Diluting, For Clinical Use
Quantity
19 EVOware Sample Tracking Software Add-on
Distribution
Nationwide distribution to CA, MA, NC, OH, TX, and VA. Worldwide distribution to Australia, Canada, China, EU, and Japan.
Recalling firm
Tecan US, Inc., Morrisville, NC

Official FDA recall record