Recall Z-3188-2026

Boston Scientific Corporation · initiated July 22, 2026

Open, Classified

Product

Boston Scientific Corporation StoneSmart Connect Console Contents: (1) Power Cord, (1) Power Adaptor, (1)DVI Cable, (1) DVI to HDMI Cable, (1) DisplayPort Cable REF: M0067933000, M006793300H0, M006793300R0, M006793300D0, M006793300Z0, M0067930000, M006793000R0

Codes / lots: All Serial Numbers: REF: M0067933000 (UDI 00191506002510, UDI (00191506021771); REF: M006793300H0 (UDI 00191506002558); REF: M006793300R0 (UDI 00191506002534), UDI (00191506021788); REF: M006793300D0 (UDI 00191506002527); REF: M006793300Z0 (UDI 00191506002541); UPN M0067933100 (UDI 00191506022549)

Reason for recall

A subset of power adapters used with the connect console may exhibit higher-than-expected resistance in the electrical grounding path. As a result, the power adaptor may fail electrical safety testing during installation.

FDA-determined cause: Nonconforming Material/Component

Action taken

On July 22, 2026, Boston Scientific Corporation issued a medical device correction to affected consignees via FedEx letter, Boston Scientific Corporation requested consignees to take the following actions: 1. Continue to use the affected devices according to the Instructions for Use. 2. Forward this notice to relevant personnel and any facilities to which the impacted device was transferred. 3. Complete and return the enclosed Acknowledgement Form. 4. Begin using the replacement power adaptor upon receipt and dispose of the affected power adaptor in accordance with local requirements.

Details

Recall number
Z-3188-2026
Initiated
July 22, 2026
Posted by FDA
September 16, 2026
Product code
FGB – Ureteroscope And Accessories, Flexible/Rigid
Quantity
982 units
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, FL, CA, OR, IL, NC, NE, UT, CO, OH, AR, NM, AL, WI, TN, IN, GA, OH, WI, KY, CT, KS, SC, IL, NE, MO, MS, DE, TX, NY, WV, MA, MT, WA, WY, NV, HI, ID, MO, ME, WV, NE, OK, MN, LA, WV, NV, PA, NH, NJ, MI, MO, AK, VA, MS, WY, ME and the countries of Australia, Finland, France, Germany, Great Britain, Greece, Ireland, Italy, Lebanon…
Recalling firm
Boston Scientific Corporation, Marlborough, MA

Official FDA recall record