Recall Z-3192-2018

Invivo Corporation · initiated August 3, 2018

Class II Terminated

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Xper Flex Cardio Physiomonitoring System Model # 453564669081 Software Version 1.6.0.0388

Codes / lots: System Serial Numbers: US1814000001, US1814000002, US1814000003, US1814000004, US1814000006, US1814000007, US1814000008, US1814000011, US1814000012, US1814000013, US1814000014, US1814000016, US1814000017, US1814000018, US1814000019, US1814000020, US1814000022, US1814000023, US1814000024, US1814000027, US1814000028, US1814000030, US1814000031, US1814000032, US1814000033, US1814000034, US1814000037, US1814000044, US1814000046, US1814000050, US1814000053, US1814000062, US1814000066, US1814000068,and US1814000072

Reason for recall

Released with an incorrect, older firmware version installed. May result in intermittent loss of ECG monitoring, inaccurate ECG amplitude display and or auxiliary output, and inaccurate heart rate (HR) displayed due to QRS detection fault.

FDA-determined cause: Software Manufacturing/Software Deployment

Action taken

Philips is asking customers to follow the Action to be Taken by Customer/User section of the August 3, 2018, Field Safety Notice: The Field Safety Notice informs customers of the issue, identifies details of the units affected, gives instructions on actions to be taken by the customer and identifies what action Philips plans to take to remedy the issue. Philips will replace all affected devices. Users should discontinue use of affected devices immediately. A Philips representative will contact you regarding your affected device.

Details

Recall number
Z-3192-2018
Initiated
August 3, 2018
Terminated
July 15, 2021
Product code
MWI – Monitor, Physiological, Patient (Without Arrhythmia Detection Or Alarms)
Quantity
38
Distribution
Distribution in the Netherlands
Recalling firm
Invivo Corporation, Orlando, FL