Recall Z-3206-2026

Howmedica Osteonics Corp. · initiated July 30, 2026

Open, Classified

Product

SCORPIO C-DOME PATELLA 73-2710 73-2910 SCORPIO U-DOME PATELLA 73-3110 73-3508 73-3708 73-3710 73-3910 Orthopaedic implant components manufactured from UHMWPE (polyethylene)

Codes / lots: 73-2710 UDI-DI 07613327033847 Lot NPPV 73-2910 UDI-DI 07613327033854 Lot 14AGFD 73-3110 UDI-DI 07613327033861 Lots 1L45 4JM5 7EL5 AME3 MK74 NTV7 PYHM RE7J VEWJ 73-3508 UDI-DI 07613327033885 Lots HMM5 NLRE 73-3708 UDI-DI 07613327033892 Lots 3467 3D8D PMRHX1 T6YJ T9ME VWDA YAM7 73-3710 UDI-DI 07613327033908 Lots 3W25 DLYP 73-3910 UDI-DI 07613327033915 Lots 14NEYA 1XED 9Y3T D1M9 D2YP EX8T KRT3 KRT3X1 T325 V8X8 X6YE

Reason for recall

Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.

FDA-determined cause: Process change control

Action taken

Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.

Details

Recall number
Z-3206-2026
Initiated
July 30, 2026
Posted by FDA
September 16, 2026
Product code
NPJ – Prosthesis, Knee Patellofemorotibial, Partial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Quantity
12,156 US
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
Recalling firm
Howmedica Osteonics Corp., Mahwah, NJ

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3195-2026 AXLE PIN ASSY LARGE ELBOW COBALT CHROME Catalog No. C-M112-2-500 Orthopaedic implant…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3196-2026 OMNIFIT SERIES I ACET. INSERT 10 22mm ID Catalog No. 2001-2246 Orthopaedic implant…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3197-2026 ALL POLY CONSTRAINED INSERT 54MM 69-2854 NRG KNEE CR 82-2-0512 82-2-0518 82-2-0912 NRG…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3198-2026 TRIATHLON HINGE INSERT SIZE 4 11MM 5612-P-411 TRIATHLON HINGE INSERT SIZE 4 13MM…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3199-2026 ABG I ACETABULAR INSERT HOODED 28mm ID X 48mm C-M079-2-801 ARC2F ACETABULAR INSERT…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3200-2026 DURACON XL ANGLED TIB INS 11/14MM C-M064-1-300 SELF-PRESSURIZING GLENOID SIZE 8…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3201-2026 ACETABULAR CUP UHMWPE 61-3258 ALL POLY CONSTRAINED INSERT 44MM 69-2244 ALL POLY…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3202-2026 BEARING AND AXLE BUSHING ASSY STD ELBOW C-M113-1-301 KRH REV TIB INS 16mm THK .640 DIA…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3203-2026 SCORPIO RECESSED METAL BACKED PATELLA 3045-0026 3045-0028 3045-0030 3045-0032…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3204-2026 TRIDENT 0 POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified
Z-3205-2026 SCORPIO + MOBILE BEARING PCS 78-5-0910 SYSTEM ONE BIPOLAR 40MM 6074-4022 SYSTEM ONE…Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged… 07/30/2026Open, Classified