Recall Z-3213-2024

Centinel Spine, Inc. · initiated July 18, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5

Codes / lots: Lot Code: UDI-DI 00843193113856; Lot 2024-0482

Reason for recall

During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.

FDA-determined cause: Process control

Action taken

Customers were contacted via telephone beginning 7/18/24. The Centinel Quality team has identified a potential nonconformance on three specific lots in the prodisc c product family. We are requesting the return of the inventory that is assigned to you. We will provide replacement inventory proactively to support the business. Thank you for your support. In addition, an URGENT - MEDICAL DEVICE Recall Instructions for PRODISC C IMPLANTS notification letter was sent to the distributor. Required Action: Please retrieve the identified devices and return the affected devices to Centinel Spine Depth of Recall: This recall should be carried out to the hospital level, including Distributors and Consigned locations Please acknowledge that you have read the email and enable the Return Receipt. This is necessary to ensure traceability as part of the recall process. If you have any questions, please contact Jessica Staub 484.887.8803.

Details

Recall number
Z-3213-2024
Initiated
July 18, 2024
Posted by FDA
September 24, 2024
Product code
MJO – Prosthesis, Intervertebral Disc
Quantity
6 units
Distribution
US Nationwide distribution.
Recalling firm
Centinel Spine, Inc., West Chester, PA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3212-2024 Class IIprodisc C SK, Total cervical disc replacement, medical device, Model/Catalog Number: PDSM5During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration… 07/18/2024Open, Classified