Recall Z-3213-2024
Centinel Spine, Inc. · initiated July 18, 2024
Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature
Product
prodisc C SK, Total cervical disc replacement, Model/Catalog Number: PDSM5
Codes / lots: Lot Code: UDI-DI 00843193113856; Lot 2024-0482
Reason for recall
During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration is considered a foreign contaminate that may not have been fully evaluated in the manufacturing validations.
FDA-determined cause: Process control
Action taken
Customers were contacted via telephone beginning 7/18/24. The Centinel Quality team has identified a potential nonconformance on three specific lots in the prodisc c product family. We are requesting the return of the inventory that is assigned to you. We will provide replacement inventory proactively to support the business. Thank you for your support. In addition, an URGENT - MEDICAL DEVICE Recall Instructions for PRODISC C IMPLANTS notification letter was sent to the distributor. Required Action: Please retrieve the identified devices and return the affected devices to Centinel Spine Depth of Recall: This recall should be carried out to the hospital level, including Distributors and Consigned locations Please acknowledge that you have read the email and enable the Return Receipt. This is necessary to ensure traceability as part of the recall process. If you have any questions, please contact Jessica Staub 484.887.8803.
Details
- Recall number
- Z-3213-2024
- Initiated
- July 18, 2024
- Posted by FDA
- September 24, 2024
- Product code
- MJO – Prosthesis, Intervertebral Disc
- Quantity
- 6 units
- Distribution
- US Nationwide distribution.
- Recalling firm
- Centinel Spine, Inc., West Chester, PA
Other products in the same recall event
| Recall | Product | Initiated |
|---|---|---|
| Z-3212-2024 Class II | prodisc C SK, Total cervical disc replacement, medical device, Model/Catalog Number: PDSM5During the assembly of Prodisc C, a blue discoloration was observed on a Prodisc C inlay. The discoloration… | 07/18/2024Open, Classified |