Recall Z-3223-2026

Ventec Life Systems, Inc. · initiated July 28, 2026

Open, Classified

Product

REF: HF2POSTD HomeFill Oxygen System HomeFill cylinder D with CGA870 Compatible Post Valve

Codes / lots: UDI:00850018761376/ Serial #s: 25EKMA00BAC 25EKMA009L3 25EKMA009CM 25EKMA009J8 25GKMA00077 25GKMA0007B 25GKMA0009Z 25GKMA0004U 25GKMA000CS 25GKMA000CR 25GKMA000CX 25GKMA000D0 25GKMA000LL 25GKMA000N1 25GKMA000UD 25GKMA000D1 25GKMA000TN 25GKMA000D2 25GKMA000UC 25GKMA000LK 25GKMA000UA 25GKMA000UB 25GKMA000UE 25GKMA000CY 25GKMA000UF

Reason for recall

Labeling; Oxygen Cylinders do not contain required medical gas labeling

FDA-determined cause: Process control

Action taken

On July 28, 2026, Ventec Life Systems Inc. issued a Urgent: Medical Device Recall notification to affected consignees via: certified mail. Ventec ask consignees to take the following actions: 1. Immediately discontinue use of affected HomeFill Oxygen Cylinders. 2. Inspect current stock, quarantine/segregate any unused affected products, remove affected product from clinical service and segregate to prevent further use until the enclosed label correction has been completed. 3. Use an unaffected oxygen cylinder for patient support. 4. If you have distributed or transferred affected product to other facilities, caregivers, or end users, you must promptly notify them of this recall and provide a copy of this notification. 5.. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health WITHIN 7 DAYS OF RECEIPT. 6. Upon receipt of the Customer Response Form, React Health will arrange for shipment of the label and instructions to apply the label to affected cylinders along with an update confirmation letter. 7. Update the affected product and return the completed confirmation letter to React Health. Contact Customer Support at 863-226 -6285 to discuss completing the form, if necessary.

Details

Recall number
Z-3223-2026
Initiated
July 28, 2026
Posted by FDA
September 18, 2026
Product code
ECX – Cylinder, Compressed Gas, And Valve
Quantity
25 units
Distribution
US Nationwide distribution in the states of FL, IA, OH, NE, TN.
Recalling firm
Ventec Life Systems, Inc., Bothell, WA

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3221-2026 HF2P0ST9 HomeFill Oxygen System HomeFill Cylinder M9 with CGA870 Compatible Post ValveLabeling; Oxygen Cylinders do not contain required medical gas labeling 07/28/2026Open, Classified
Z-3222-2026 REF:HF2PC9 HomeFill Oxygen System, HomeFill Cylinder M9 with Integrated ConserverLabeling; Oxygen Cylinders do not contain required medical gas labeling 07/28/2026Open, Classified