Recall Z-3229-2024

American Contract Systems, Inc. · initiated August 7, 2024

Class II Open, Classified

Class II – products that might cause a temporary health problem or pose a slight threat of a serious nature

Product

Arthroscopy Pack, MSKA45A; Medical convenience kit

Codes / lots: UDI-DI: 00191072200938; Lot: 2405153; Exp: 05/15/2025

Reason for recall

ACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one of the controlled environment areas. As a result, ACS is unable to confirm product sterilization assurance requirements were met.

FDA-determined cause: Under Investigation by firm

Action taken

Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " -Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Owens & Minor issued an URGENT: MEDICAL DEVICE FIELD CORRECTION to its consignees on 08/07/2024 via email. The notice explained the issue, potential risk to health, and requested the following actions be taken: " Please review your inventory, and segregate and quarantine all affected Trays. If any of the affected trays identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. - Discard all affected kits and document on the enclosed Response Form."

Details

Recall number
Z-3229-2024
Initiated
August 7, 2024
Posted by FDA
September 24, 2024
Product code
OJH – Orthopedic Tray
Quantity
48 units
Distribution
US Nationwide distribution in the states of MO, MN, MA, OH, NE.
Recalling firm
American Contract Systems, Inc., Kansas City, MO

Official FDA recall record

Other products in the same recall event

RecallProductInitiated
Z-3226-2024 Class IIPACEMAKER PACK, SLCV01J; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3227-2024 Class IIEndovascular AAA Pack, ESED52B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3228-2024 Class IIEndovascular AAA Pack, ESED52B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3230-2024 Class IIOpen Shoulder, LWOS39L; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3231-2024 Class IIDay Surgery General Pack, FHGE18T; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3232-2024 Class IIShoulder, SESH18I; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3233-2024 Class IIBilateral Tubal Ligation Pack, FHBT3T; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3234-2024 Class IIParacentesis Thoracente, SLPT76H; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3235-2024 Class IIIR Procedure Pack, CEAT19W; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3236-2024 Class IINeuro Vascular, ANKV91A; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3237-2024 Class IIHand Pack, AKHD97C; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3238-2024 Class IIVaginal Delivery, ASDV25B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3239-2024 Class IICath Lab Pack, CECL02P; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3240-2024 Class IIEssentia MAJ Laparotomy Std Pack, ESLT53B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3241-2024 Class IIENT, COEN27N; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3242-2024 Class IINasal Sinus Pack, MTSN26B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3243-2024 Class IICysto, FTCY03L; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified
Z-3244-2024 Class IIPacemaker, COPM11B; Medical convenience kitACS identified that due to an inoperable chart recorder, humidity readings were out-of-specification in one… 08/07/2024Open, Classified